Radiation dose reduction in direct digital panoramic radiography
- PMID: 18448296
- DOI: 10.1016/j.ejrad.2008.03.018
Radiation dose reduction in direct digital panoramic radiography
Abstract
Objectives: (a) To measure the absorbed radiation doses at 16 anatomical sites of a Rando phantom and (b) to calculate the effective doses including and excluding the salivary gland doses in panoramic radiography using a conventional and a digital panoramic device.
Study design: Thermoluminescent dosimeters (TLD-100) were placed at 16 sites in a Rando phantom, using a conventional, Planmeca Promax and a digital, Planmeca PM2002CC Proline 2000 (Planmeca Oy, 00880 Helsinki, Finland) panoramic device for panoramic radiography. During conventional radiography the selected exposure settings were 66 kVp, 6 mA and 16s, while during digital radiography two combinations were selected 60 kVp, 4 mA, 18 s and 66 kVp, 8 mA, 18s with and without image processing function. The dosimeters were annealed in a PTW-TLDO Harshaw oven. TLD energy response was studied using RQN beam narrow series at GAEC's Secondary Standard Calibration Laboratory. The reader used was a Harshaw, 4500. Effective dose was estimated according to ICRP(60) report (E(ICRP60)). An additional estimation of the effective dose was accomplished including the doses of the salivary glands (E(SAL)). A Wilcoxon signed ranks test was used for statistical analysis.
Results: The effective dose, according to ICRP report (E(ICRP60)) in conventional panoramic radiography was 17 microSv and E(SAL) was 26 microSv. The respective values in digital panoramic radiography were E(ICRP60)=23 microSv and E(SAL)=38 microSv; while using the lowest possible radiographic settings E(ICRP60) was 8 microSv and E(SAL) was 12 microSv.
Conclusions: The effective dose reduction in digital panoramic radiography can be achieved, if the lowest possible radiographic settings are used.
Similar articles
-
Radiation doses of indirect and direct digital cephalometric radiography.Br Dent J. 2004 Aug 14;197(3):149-52; discussion 140. doi: 10.1038/sj.bdj.4811532. Br Dent J. 2004. PMID: 15311250
-
Effective dose from radiation absorbed during a panoramic examination with a new generation machine.Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2000 Feb;89(2):236-43. doi: 10.1067/moe.2000.103526. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2000. PMID: 10673663
-
Effective dosages for recording Veraviewepocs dental panoramic images: analog film, digital, and panoramic scout for CBCT.Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2008 Oct;106(4):571-7. doi: 10.1016/j.tripleo.2008.03.031. Epub 2008 Jul 7. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2008. PMID: 18602314
-
The current status of panoramic radiography.Int Dent J. 1987 Mar;37(1):31-7. Int Dent J. 1987. PMID: 3294598 Review.
-
Update on the biological effects of ionizing radiation, relative dose factors and radiation hygiene.Aust Dent J. 2012 Mar;57 Suppl 1:2-8. doi: 10.1111/j.1834-7819.2011.01665.x. Aust Dent J. 2012. PMID: 22376091 Review.
Cited by
-
The suitability of panoramic radiographs for clinical decision making regarding root angulation compared to cone-beam computed tomography.BMC Med Imaging. 2021 May 24;21(1):89. doi: 10.1186/s12880-021-00619-y. BMC Med Imaging. 2021. PMID: 34030659 Free PMC article.
-
Skin entrance dose with and without lead apron in digital panoramic radiography for selected sensitive body regions.Clin Oral Investig. 2017 May;21(4):1327-1333. doi: 10.1007/s00784-016-1886-0. Epub 2016 Jun 21. Clin Oral Investig. 2017. PMID: 27324474
-
Modern 3D cephalometry in pediatric orthodontics-downsizing the FOV and development of a new 3D cephalometric analysis within a minimized large FOV for dose reduction.Clin Oral Investig. 2021 Jul;25(7):4651-4670. doi: 10.1007/s00784-021-03779-x. Epub 2021 Jan 25. Clin Oral Investig. 2021. PMID: 33492515
-
Characterization of optically stimulated luminescence dosemeters to measure organ doses in diagnostic radiology.Dentomaxillofac Radiol. 2012 Mar;41(3):211-6. doi: 10.1259/dmfr/98708146. Epub 2011 Nov 24. Dentomaxillofac Radiol. 2012. PMID: 22116136 Free PMC article.
-
Image quality assessment in panoramic dental radiography: a comparative study between conventional and digital systems.Quant Imaging Med Surg. 2013 Feb;3(1):43-8. doi: 10.3978/j.issn.2223-4292.2013.02.07. Quant Imaging Med Surg. 2013. PMID: 23483085 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources