Radiation dose reduction in direct digital panoramic radiography
- PMID: 18448296
- DOI: 10.1016/j.ejrad.2008.03.018
Radiation dose reduction in direct digital panoramic radiography
Abstract
Objectives: (a) To measure the absorbed radiation doses at 16 anatomical sites of a Rando phantom and (b) to calculate the effective doses including and excluding the salivary gland doses in panoramic radiography using a conventional and a digital panoramic device.
Study design: Thermoluminescent dosimeters (TLD-100) were placed at 16 sites in a Rando phantom, using a conventional, Planmeca Promax and a digital, Planmeca PM2002CC Proline 2000 (Planmeca Oy, 00880 Helsinki, Finland) panoramic device for panoramic radiography. During conventional radiography the selected exposure settings were 66 kVp, 6 mA and 16s, while during digital radiography two combinations were selected 60 kVp, 4 mA, 18 s and 66 kVp, 8 mA, 18s with and without image processing function. The dosimeters were annealed in a PTW-TLDO Harshaw oven. TLD energy response was studied using RQN beam narrow series at GAEC's Secondary Standard Calibration Laboratory. The reader used was a Harshaw, 4500. Effective dose was estimated according to ICRP(60) report (E(ICRP60)). An additional estimation of the effective dose was accomplished including the doses of the salivary glands (E(SAL)). A Wilcoxon signed ranks test was used for statistical analysis.
Results: The effective dose, according to ICRP report (E(ICRP60)) in conventional panoramic radiography was 17 microSv and E(SAL) was 26 microSv. The respective values in digital panoramic radiography were E(ICRP60)=23 microSv and E(SAL)=38 microSv; while using the lowest possible radiographic settings E(ICRP60) was 8 microSv and E(SAL) was 12 microSv.
Conclusions: The effective dose reduction in digital panoramic radiography can be achieved, if the lowest possible radiographic settings are used.