Design and instrumentation of new devices for performing appendectomy at colonoscopy (with video)
- PMID: 18455170
- DOI: 10.1016/j.gie.2008.02.024
Design and instrumentation of new devices for performing appendectomy at colonoscopy (with video)
Abstract
Background: There is increasing interest in natural orifice surgery (NOS). Because the lumen of the appendix is connected to the cecum, a minimally invasive method for removing the appendix by colonoscopy may be feasible.
Objectives: Our purpose was to design, develop, and test new devices for inverting and removing the appendix by colonoscopy.
Design: Prospective prototype development program.
Setting: University-based study in 25 colons from adult human cadavers.
Interventions and methods: Various prototypes were evaluated by inserting them into the appendiceal orifice to its luminal tip, with the intent to invert the appendix in a controlled fashion into the lumen of the cecum. The advantage of using a tubular structure as a counterforce to aid inversion of the appendix was evaluated. When inversion was incomplete, the growing tissue strain was relieved by endoluminal incision of the mesenteric side of the appendix. Closure methods with endoloops or ligating loops were studied. Appendiceal resection was completed by snare diathermy, leaving an inverted intraluminal stump.
Main outcome measurement: Ability to invert the appendix into the cecum.
Results: The mean appendix length and luminal diameter were 84 +/- 23 mm and 4.9 +/- 1.2 mm, respectively. It was possible to advance various types of inversion devices to the tip of the appendiceal lumen. Partial inversion of the appendix was successful in 22 of 25 tests. Mesenteric tissue tension, tissue volume, and device slippage were the main reasons for incomplete inversion. The complete inversion was achieved with a combination of vacuum, tip grip, counterforce at the appendix base, and eventually endoluminal incision.
Conclusions: The inversion of the human appendix by colonoscopy seems feasible and may be an alternative approach to conventional appendectomy.
Similar articles
-
Inversion-ligation appendectomy for incidental appendectomy.Am J Surg. 1991 Apr;161(4):483-4. doi: 10.1016/0002-9610(91)91118-3. Am J Surg. 1991. PMID: 2035769
-
Colonoscopic diagnosis of mucocele of the appendix.Gastrointest Endosc. 2005 Sep;62(3):452-6. doi: 10.1016/j.gie.2005.04.018. Gastrointest Endosc. 2005. PMID: 16111974
-
A motor-driven single-use colonoscope controlled with a hand-held device: a feasibility study in volunteers.Gastrointest Endosc. 2008 Jun;67(7):1139-46. doi: 10.1016/j.gie.2007.10.065. Epub 2008 Mar 19. Gastrointest Endosc. 2008. PMID: 18355823
-
Renewed attention for overtube-assisted colonoscopy to prevent incomplete endoscopic examination of the colon.Dis Colon Rectum. 2013 Aug;56(8):1013-8. doi: 10.1097/DCR.0b013e3182962636. Dis Colon Rectum. 2013. PMID: 23838871 Review.
-
Management of ingested foreign bodies within the appendix: a case report with review of the literature.Am J Gastroenterol. 1997 Dec;92(12):2295-8. Am J Gastroenterol. 1997. PMID: 9399774 Review.
Cited by
-
Ilececum: A Comprehensive Review.Can J Gastroenterol Hepatol. 2019 Feb 3;2019:1451835. doi: 10.1155/2019/1451835. eCollection 2019. Can J Gastroenterol Hepatol. 2019. PMID: 30854348 Free PMC article. Review.
-
Endoluminal and transluminal surgery: current status and future possibilities.Surg Endosc. 2006 Aug;20(8):1179-92. doi: 10.1007/s00464-005-0711-7. Epub 2006 Jul 24. Surg Endosc. 2006. PMID: 16865615 Review.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical