Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Comparative Study
. 2008 Jun;43(6):343-8.
doi: 10.1097/RLI.0b013e318168f705.

Comparison of radiologist and CAD performance in the detection of CT-confirmed subtle pulmonary nodules on digital chest radiographs

Affiliations
Comparative Study

Comparison of radiologist and CAD performance in the detection of CT-confirmed subtle pulmonary nodules on digital chest radiographs

Thorsten Alexander Bley et al. Invest Radiol. 2008 Jun.

Abstract

Objectives: Detection of subtle pulmonary nodules on digital radiography is a challenging task for radiologists. The aim of this study was to evaluate the performance of a newly approved computer aided detection (CAD) system.

Materials and methods: The sensitivity of 3 radiologists and of a CAD system for the detection of pulmonary nodules from 5 to 15 mm in size on digital chest radiography of 117 patients was compared. The reference standard was established by consensus reading of computed tomography scans by 2 experienced radiologists. Computed tomography scans and chest radiographs were performed within 4 weeks. Sixty-six pulmonary nodules from 42 patients, with a mean nodule diameter of 7.5 mm (standard deviation: 2.2 mm), were included in the statistical analysis. Seventy-five of the 117 patients did not have nodules from 5 to 15 mm of size.

Results: Two hundred and eighty-eight false-positive detections of the CAD system were found with an average of 2.5 false-positives per image. Sensitivity of the CAD system was 39.4% (95% confidence interval: 11.8%), when compared with 18.2% to 30.3% (95% confidence interval 9.3% to 11.1%) of the 3 radiologists. Substantial agreement for nodule detection ([kappa]N: 0.64-0.73) was found among the 3 radiologists, whereas only moderate agreement was found between the radiologists and the CAD performance ([kappa]N: 0.45-0.52).

Conclusions: The CAD system's diagnostic sensitivity in detecting pulmonary nodules of 5 to 15 mm of size was superior to the 1 of radiologists. The CAD system may be used for assisting the radiologist in the detection of lung nodules on digital chest radiographs.

PubMed Disclaimer