Comparative Study
. 2008 Sep;34(9):1718-23.
doi: 10.1007/s00134-008-1126-4. Epub 2008 Apr 30.

Comparison of lung tissue concentrations of nebulized ceftazidime in ventilated piglets: ultrasonic versus vibrating plate nebulizers

Affiliations
Comparative Study

Comparison of lung tissue concentrations of nebulized ceftazidime in ventilated piglets: ultrasonic versus vibrating plate nebulizers

Fabio Ferrari et al. Intensive Care Med. 2008 Sep.

Abstract

Objective: To compare the efficiency of an Aeroneb Pro vibrating plate and an Atomisor MegaHertz ultrasonic nebulizer for providing ceftazidime distal lung deposition.

Design: In vitro experiments. One gram of cetazidime was nebulized in respiratory circuits and mass median aerodynamic diameter of particles generated by ultrasonic and vibrating plate nebulizers was compared using a laser velocimeter. In vivo experiments. Lung tissue concentrations and extrapulmonary depositions were measured in ten anesthetized ventilated piglets with healthy lungs that received 1 g of ceftazidime by nebulization with either an ultrasonic (n = 5), or a vibrating plate (n = 5) nebulizer.

Setting: A two-bed Experimental Intensive Care Unit of a University School of Medicine.

Intervention: Following sacrifice, 5 subpleural specimens were sampled in dependent and nondependent lung regions for measuring ceftazidime lung tissue concentrations by high-performance liquid chromatography.

Measurements and results: Mass median aerodynamic diameters generated by both nebulizers were similar with more than 95% of the particles between 0.5 and 5 microm. Lung tissue concentrations were 553 +/- 123 [95% confidence interval: 514-638] microg g(-1) using ultrasonic nebulizer, and 452 +/- 172 [95% confidence interval: 376-528] microg g(-1) using vibrating plate nebulizers (NS). Extrapulmonary depositions were, respectively, of 38 +/- 5% (ultrasonic) and 34 +/- 4% (vibrating plate) (NS).

Conclusions: Vibrating plate nebulizer is comparable to ultrasonic nebulizers for ceftazidime nebulization. It may represent a new attractive technology for inhaled antibiotic therapy.

PubMed Disclaimer

Similar articles

Cited by

References

    1. Eur Respir J. 1997 Apr;10(4):905-9 - PubMed
    1. Pediatr Res. 2005 Jul;58(1):10-4 - PubMed
    1. Anesthesiology. 2005 May;102(5):995-1000 - PubMed
    1. J Aerosol Med. 1997 Spring;10(1):13-23 - PubMed
    1. Am J Respir Crit Care Med. 2002 Nov 15;166(10):1375-81 - PubMed

Publication types

LinkOut - more resources