Comparison of lung tissue concentrations of nebulized ceftazidime in ventilated piglets: ultrasonic versus vibrating plate nebulizers
- PMID: 18542922
- DOI: 10.1007/s00134-008-1126-4
Comparison of lung tissue concentrations of nebulized ceftazidime in ventilated piglets: ultrasonic versus vibrating plate nebulizers
Abstract
Objective: To compare the efficiency of an Aeroneb Pro vibrating plate and an Atomisor MegaHertz ultrasonic nebulizer for providing ceftazidime distal lung deposition.
Design: In vitro experiments. One gram of cetazidime was nebulized in respiratory circuits and mass median aerodynamic diameter of particles generated by ultrasonic and vibrating plate nebulizers was compared using a laser velocimeter. In vivo experiments. Lung tissue concentrations and extrapulmonary depositions were measured in ten anesthetized ventilated piglets with healthy lungs that received 1 g of ceftazidime by nebulization with either an ultrasonic (n = 5), or a vibrating plate (n = 5) nebulizer.
Setting: A two-bed Experimental Intensive Care Unit of a University School of Medicine.
Intervention: Following sacrifice, 5 subpleural specimens were sampled in dependent and nondependent lung regions for measuring ceftazidime lung tissue concentrations by high-performance liquid chromatography.
Measurements and results: Mass median aerodynamic diameters generated by both nebulizers were similar with more than 95% of the particles between 0.5 and 5 microm. Lung tissue concentrations were 553 +/- 123 [95% confidence interval: 514-638] microg g(-1) using ultrasonic nebulizer, and 452 +/- 172 [95% confidence interval: 376-528] microg g(-1) using vibrating plate nebulizers (NS). Extrapulmonary depositions were, respectively, of 38 +/- 5% (ultrasonic) and 34 +/- 4% (vibrating plate) (NS).
Conclusions: Vibrating plate nebulizer is comparable to ultrasonic nebulizers for ceftazidime nebulization. It may represent a new attractive technology for inhaled antibiotic therapy.
Similar articles
-
Intravenous versus nebulized ceftazidime in ventilated piglets with and without experimental bronchopneumonia: comparative effects of helium and nitrogen.Anesthesiology. 2005 May;102(5):995-1000. doi: 10.1097/00000542-200505000-00019. Anesthesiology. 2005. PMID: 15851887
-
Nebulized ceftazidime in experimental pneumonia caused by partially resistant Pseudomonas aeruginosa.Intensive Care Med. 2009 Oct;35(10):1792-800. doi: 10.1007/s00134-009-1605-2. Epub 2009 Aug 4. Intensive Care Med. 2009. PMID: 19652947
-
Albuterol delivery in an in vitro pediatric ventilator lung model: comparison of jet, ultrasonic, and mesh nebulizers.Pediatr Crit Care Med. 2013 Feb;14(2):e98-102. doi: 10.1097/PCC.0b013e3182712783. Pediatr Crit Care Med. 2013. PMID: 23287904
-
A review of the technical aspects of drug nebulization.Pharm World Sci. 2000 Jun;22(3):75-81. doi: 10.1023/a:1008786600530. Pharm World Sci. 2000. PMID: 11028259 Review.
-
Design principles of liquid nebulization devices currently in use.Respir Care. 2002 Nov;47(11):1257-75; discussion 1275-8. Respir Care. 2002. PMID: 12425742 Review.
Cited by
-
Optimizing Antimicrobial Drug Dosing in Critically Ill Patients.Microorganisms. 2021 Jun 28;9(7):1401. doi: 10.3390/microorganisms9071401. Microorganisms. 2021. PMID: 34203510 Free PMC article. Review.
-
Aerosol delivery with two ventilation modes during mechanical ventilation: a randomized study.Ann Intensive Care. 2016 Dec;6(1):73. doi: 10.1186/s13613-016-0169-x. Epub 2016 Jul 22. Ann Intensive Care. 2016. PMID: 27447788 Free PMC article.
-
Aerosol delivery during invasive mechanical ventilation: a systematic review.Crit Care. 2017 Oct 21;21(1):264. doi: 10.1186/s13054-017-1844-5. Crit Care. 2017. PMID: 29058607 Free PMC article.
-
Utilizing a three-dimensional printed additional nebulization chamber for colistin nebulization with a large diluent volume through a vibrating mesh nebulizer: proof of concept.Eur Clin Respir J. 2025 Jul 11;12(1):2531615. doi: 10.1080/20018525.2025.2531615. eCollection 2025. Eur Clin Respir J. 2025. PMID: 40661897 Free PMC article.
-
Adapting the Aerogen Mesh Nebulizer for Dried Aerosol Exposures Using the PreciseInhale Platform.J Aerosol Med Pulm Drug Deliv. 2020 Apr;33(2):116-126. doi: 10.1089/jamp.2019.1554. Epub 2019 Oct 15. J Aerosol Med Pulm Drug Deliv. 2020. PMID: 31613690 Free PMC article.
References
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical
Miscellaneous