Comparative Study
. 2008 Aug;46(8):2613-9.
doi: 10.1128/JCM.02237-07. Epub 2008 Jun 11.

Gram stain-specific-probe-based real-time PCR for diagnosis and discrimination of bacterial neonatal sepsis

Affiliations
Comparative Study

Gram stain-specific-probe-based real-time PCR for diagnosis and discrimination of bacterial neonatal sepsis

Yi-Dong Wu et al. J Clin Microbiol. 2008 Aug.

Abstract

Sepsis is a serious disease with high mortality in newborns. It is very important to have a convenient and accurate method for pathogenic diagnosis of neonatal sepsis. We developed a method of simultaneous detection and Gram classification of clinically relevant bacterial pathogens causing sepsis directly from blood samples with Gram stain-specific-probe-based real-time PCR (GSPBRT-PCR). With GSPBRT-PCR, 53 clinically important strains representing 25 gram-positive and 28 gram-negative bacterial species were identified correctly with the corresponding Gram probe. The limits of the GSPBRT-PCR assay in serial dilutions of the bacteria revealed that Staphylococcus aureus could be detected at concentrations of 3 CFU per PCR and Escherichia coli at concentrations as low as 1 CFU per PCR. The GSPBRT-PCR assay was further evaluated on 600 blood specimens from patients with suspicion of neonatal sepsis and compared to the results obtained from blood cultures. The positive rate of the GSPBRT-PCR array was 50/600 (8.33%), significantly higher than that of blood culture (34/600; 5.67%) (P = 0.00003). When blood culture was used as a control, the sensitivity of GSPBRT-PCR was 100%, the specificity was 97.17%, and the index of accurate diagnosis was 0.972. This study suggests that GSPBRT-PCR is very useful for the rapid and accurate diagnosis of bacterial infection and that it can have an important impact on the current inappropriate and unnecessary use of antibiotics in the treatment of newborns.

PubMed Disclaimer

Figures

FIG. 1.
FIG. 1.
Comparison of the standard curves of the Gram stain-specific probes from serial dilutions of bacteria. (A) Standard curve of G+ probe for S. aureus. (B) Standard curve of G probe for E. coli.

References

    1. Anand, A. R., H. N. Madhavan, and K. L. Therese. 2000. Use of polymerase chain reaction (PCR) and DNA probe hybridization to determine the Gram reaction of the infecting bacterium in the intraocular fluids of patients with endophthalmitis. J. Infect. 41221-226. - PubMed
    1. Anonymous. 1995. Case records of the Massachusetts General Hospital. Weekly clinicopathological exercises. Case 11-1995. A 39-year-old man with chronic renal failure, aortic regurgitation, and a calcified mass around the aortic root. N. Engl. J. Med. 3321015-1022. - PubMed
    1. Bergeron, M. G., D. Ke, C. Menard, F. J. Picard, M. Gagnon, M. Bernier, M. Ouellette, P. H. Roy, S. Marcoux, and W. D. Fraser. 2000. Rapid detection of group B streptococci in pregnant women at delivery. N. Engl. J. Med. 343175-179. - PubMed
    1. Carroll, N. M., P. Adamson, and N. Okhravi. 1999. Elimination of bacterial DNA from Taq DNA polymerases by restriction endonuclease digestion. J. Clin. Microbiol. 373402-3404. - PMC - PubMed
    1. Carroll, N. M., E. E. Jaeger, S. Choudhury, A. A. Dunlop, M. M. Matheson, P. Adamson, N. Okhravi, and S. Lightman. 2000. Detection of and discrimination between gram-positive and gram-negative bacteria in intraocular samples by using nested PCR. J. Clin. Microbiol. 381753-1757. - PMC - PubMed

Publication types