Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2008 May;35(5):1950-8.
doi: 10.1118/1.2903436.

Dedicated breast computed tomography: volume image denoising via a partial-diffusion equation based technique

Affiliations

Dedicated breast computed tomography: volume image denoising via a partial-diffusion equation based technique

Jessie Q Xia et al. Med Phys. 2008 May.

Abstract

Dedicated breast computed tomography (CT) imaging possesses the potential for improved lesion detection over conventional mammograms, especially for women with dense breasts. The breast CT images are acquired with a glandular dose comparable to that of standard two-view mammography for a single breast. Due to dose constraints, the reconstructed volume has a non-negligible quantum noise when thin section CT slices are visualized. It is thus desirable to reduce noise in the reconstructed breast volume without loss of spatial resolution. In this study, partial diffusion equation (PDE) based denoising techniques specifically for breast CT were applied at different steps along the reconstruction process and it was found that denoising performed better when applied to the projection data rather than reconstructed data. Simulation results from the contrast detail phantom show that the PDE technique outperforms Wiener denoising as well as adaptive trimmed mean filter. The PDE technique increases its performance advantage relative to Wiener techniques when the photon fluence is reduced. With the PDE technique, the sensitivity for lesion detection using the contrast detail phantom drops by less than 7% when the dose is cut down to 40% of the two-view mammography. For subjective evaluation, the PDE technique was applied to two human subject breast data sets acquired on a prototype breast CT system. The denoised images had appealing visual characteristics with much lower noise levels and improved tissue textures while maintaining sharpness of the original reconstructed volume.

PubMed Disclaimer

Figures

Figure 1
Figure 1
Illustration of a dedicated breast CT system. The x-ray tube and flat-panel detector rotate together around the breast, which is the only region to be illuminated.
Figure 2
Figure 2
Illustration of possible steps where image-denoising module may be applied with respect to reconstruction module for dedicated breast CT data. In this study, the denoising techniques are applied at steps 2–4.
Figure 3
Figure 3
One-dimensional line integral profiles across the breast on a projection image. The dashed and continuous plots correspond to noise free case and the case with I0=2.0e3, respectively. The variance of line integral is larger at the center of breast region and gets lower toward the periphery.
Figure 4
Figure 4
Step comparison at I0=2.5e4. (a) is a coronal slice of simulated breast with contrast detail phantoms, and (b) shows the same slice with added noise to the projection images. After noise removal, (d) PDEtomo applied prior to reconstruction generates better images than (c) PDE3Dpost after reconstruction.
Figure 5
Figure 5
Step comparison at I0=1.0e4. (b) is noisier than Fig. 4b. As in Fig. 4: (d) PDEtomo applied at step 2 is better than (c) PDE3Dpost applied at step 4.
Figure 6
Figure 6
Noise-resolution plot for PDEtomo, ATM, and Wiener filters at I0=2.5e4. For a high contrast test object, PDEtomo and ATM have a lower noise level and higher resolution than Wiener filtering. ATM is the best among the three based on the noise-resolution plot.
Figure 7
Figure 7
One-dimensional profiles of the reconstruction through the center of the high contrast object using ATM filter and PDEtomo. It is shown that by using ATM filter there is a cupping effect at the lesion center, which is absent from PDEtomo and Wiener processed ones. The rectangle is where the noise level is computed for noise-resolution plot.
Figure 8
Figure 8
Human subject result No. 1. Top row shows original reconstruction, coronal slices from normal breast. Bottom row shows same slices with PDEtomo denoising, resulting in remarkably reduced noise levels.
Figure 9
Figure 9
Human subject result No. 2. Top row shows original reconstruction, coronal slices from normal breast. Bottom row shows same slices with PDEtomo denoising, resulting in markedly reduced noise levels.

Similar articles

Cited by

References

    1. ACS, American Cancer Society: Cancer Facts and Figures 2008 (American Cancer Society, Atlanta, Georgia, 2008).
    1. Fletcher S. W. and Elmore J. G., “Mammographic screening for breast cancer,” N. Engl. J. Med. NEJMAG10.1056/NEJMcp021804 348, 1672–1680 (2003). - DOI - PMC - PubMed
    1. Nystrom L., Andersson I., Bjurstam N., Frisell J., Nordenskjold B., and Rutqvist L. E., “Long-term effects of mammography screening: Updated overview of the swedish randomised trials,” Lancet LANCAO10.1016/S0140-6736(02)08020-0 359, 909–919 (2002). - DOI - PubMed
    1. Jackson V. P., Hendrick R. E., Feig S. A., and Kopans D. B., “Imaging of the radiographically dense breast,” Radiology 188, 297–301 (1993). - PubMed
    1. Niklason L. T. et al., “Digital tomosynthesis in breast imaging,” Radiology 205, 399–406 (1997). - PubMed

Publication types