On the use of Monte Carlo-derived dosimetric data in the estimation of patient dose from CT examinations
- PMID: 18561678
- DOI: 10.1118/1.2896075
On the use of Monte Carlo-derived dosimetric data in the estimation of patient dose from CT examinations
Abstract
The purpose of this work was to investigate the applicability and appropriateness of Monte Carlo-derived normalized data to provide accurate estimations of patient dose from computed tomography (CT) exposures. Monte Carlo methodology and mathematical anthropomorphic phantoms were used to simulate standard patient CT examinations of the head, thorax, abdomen, and trunk performed on a multislice CT scanner. Phantoms were generated to simulate the average adult individual and two individuals with different body sizes. Normalized dose values for all radiosensitive organs and normalized effective dose values were calculated for standard axial and spiral CT examinations. Discrepancies in CT dosimetry using Monte Carlo-derived coefficients originating from the use of: (a) Conversion coefficients derived for axial CT exposures, (b) a mathematical anthropomorphic phantom of standard body size to derive conversion coefficients, and (c) data derived for a specific CT scanner to estimate patient dose from CT examinations performed on a different scanner, were separately evaluated. The percentage differences between the normalized organ dose values derived for contiguous axial scans and the corresponding values derived for spiral scans with pitch = 1 and the same total scanning length were up to 10%, while the corresponding percentage differences in normalized effective dose values were less than 0.7% for all standard CT examinations. The normalized organ dose values for standard spiral CT examinations with pitch 0.5-1.5 were found to differ from the corresponding values derived for contiguous axial scans divided by the pitch, by less than 14% while the corresponding percentage differences in normalized effective dose values were less than 1% for all standard CT examinations. Normalized effective dose values for the standard contiguous axial CT examinations derived by Monte Carlo simulation were found to considerably decrease with increasing body size of the mathematical phantom used. When the body-mass index was increased from 23.0 to 32.7 kg/m2 discrepancies in patient effective dose were up to 34%. The error in estimating effective dose from a CT exposure performed on a specific CT scanner using Monte Carlo data derived for a different CT scanner was estimated to be up to 25%. A simple method was proposed and validated for the determination of scanner-specific normalized dosimetric data from data derived from Monte Carlo simulation of a specific scanner. In conclusion, computed tomography dose index (CTDI) to effective dose conversion coefficients derived by Monte Carlo simulation of axial CT scans may provide a good approximation of corresponding coefficients applicable in helical scans. However, the use of Monte Carlo conversion coefficients for the estimation of patient dose from a CT examination involves a remarkable inaccuracy when the body size of the mathematical anthropomorphic phantom used in Monte Carlo simulation differs from the body of the patient. Therefore, separate sets of Monte Carlo dosimetric CT data shall be generated for different patient body sizes. Besides calculation of different sets of Monte Carlo data for each commercially available scanner is not necessary, since scanner specific data may be derived with acceptable accuracy from the Monte Carlo data calculated for a specific scanner appropriately modified for the different CTDI(W)/CTDI(air) ratio.
Similar articles
-
The effect of z overscanning on patient effective dose from multidetector helical computed tomography examinations.Med Phys. 2005 Jun;32(6):1621-9. doi: 10.1118/1.1924309. Med Phys. 2005. PMID: 16013721
-
A Monte Carlo based method to estimate radiation dose from multidetector CT (MDCT): cylindrical and anthropomorphic phantoms.Phys Med Biol. 2005 Sep 7;50(17):3989-4004. doi: 10.1088/0031-9155/50/17/005. Epub 2005 Aug 11. Phys Med Biol. 2005. PMID: 16177525
-
Size-specific, scanner-independent organ dose estimates in contiguous axial and helical head CT examinations.Med Phys. 2014 Dec;41(12):121909. doi: 10.1118/1.4901517. Med Phys. 2014. PMID: 25471969
-
CT Dosimetry: What Has Been Achieved and What Remains to Be Done.Invest Radiol. 2021 Jan;56(1):62-68. doi: 10.1097/RLI.0000000000000727. Invest Radiol. 2021. PMID: 32932380 Review.
-
USE OF ANTHROPOMORPHIC HETEROGENEOUS PHYSICAL PHANTOMS FOR VALIDATION OF COMPUTATIONAL DOSIMETRY OF MEDICAL PERSONNEL AND PATIENTS.Probl Radiac Med Radiobiol. 2020 Dec;25:148-176. doi: 10.33145/2304-8336-2020-25-148-176. Probl Radiac Med Radiobiol. 2020. PMID: 33361833 Review. English, Ukrainian.
Cited by
-
Reducing radiation dose to selected organs by selecting the tube start angle in MDCT helical scans: a Monte Carlo based study.Med Phys. 2009 Dec;36(12):5654-64. doi: 10.1118/1.3259773. Med Phys. 2009. PMID: 20095278 Free PMC article.
-
Accuracy of patient-specific organ dose estimates obtained using an automated image segmentation algorithm.J Med Imaging (Bellingham). 2016 Oct;3(4):043502. doi: 10.1117/1.JMI.3.4.043502. Epub 2016 Nov 29. J Med Imaging (Bellingham). 2016. PMID: 27921070 Free PMC article.
-
Evaluation of dose conversion coefficients for an eight-year-old Iranian male phantom undergoing computed tomography.Radiat Environ Biophys. 2015 Nov;54(4):465-74. doi: 10.1007/s00411-015-0607-4. Epub 2015 Jun 17. Radiat Environ Biophys. 2015. PMID: 26082027
-
Technical note: Evaluation of a V-Net autosegmentation algorithm for pediatric CT scans: Performance, generalizability, and application to patient-specific CT dosimetry.Med Phys. 2022 Apr;49(4):2342-2354. doi: 10.1002/mp.15521. Epub 2022 Feb 22. Med Phys. 2022. PMID: 35128672 Free PMC article.
-
Estimating peak skin and eye lens dose from neuroperfusion examinations: use of Monte Carlo based simulations and comparisons to CTDIvol, AAPM Report No. 111, and ImPACT dosimetry tool values.Med Phys. 2013 Sep;40(9):091901. doi: 10.1118/1.4816652. Med Phys. 2013. PMID: 24007152 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Medical