Comparative Study
. 2008 Aug;46(8):2681-5.
doi: 10.1128/JCM.00449-08. Epub 2008 Jun 25.

Multicenter evaluation of the new Vitek 2 Neisseria-Haemophilus identification card

Affiliations
Comparative Study

Multicenter evaluation of the new Vitek 2 Neisseria-Haemophilus identification card

Robert P Rennie et al. J Clin Microbiol. 2008 Aug.

Abstract

The new Neisseria-Haemophilus identification (NH) card for Vitek 2 was compared with 16S rRNA gene sequencing (16S) as the reference method for accurate identification of Neisseria spp., Haemophilus spp., and other fastidious gram-negative bacteria. Testing was performed on the Vitek 2 XL system with modified software at three clinical trial laboratories. Reproducibility was determined with nine ATCC quality control strains tested 20 times over a minimum of 10 days at all three sites. A challenge set of 30 strains with known identifications and 371 recent fresh and frozen clinical isolates were also tested. Expected positive and negative biochemical reactions were also evaluated for substrate reproducibility. All microorganisms were tested on the NH card, and all clinical and stock isolates were saved for 16S testing. All reproducibility tests yielded expected results within a 95% confidence interval. For challenge microorganisms, there was 98% overall correct identification, including 8% low discrimination, 2% incorrect identification, and 0% unidentified. For clinical strains, there was 96.5% overall correct identification, including 10.2% low discrimination, 2.7% incorrect identification, and 0.8% unidentified. The 2.7% (10/371) of clinical isolates that gave an incorrect identification consisted of 7 isolates correct to genus and 3 strains incorrect to genus. There were an additional 27 strains (primarily Neisseria species) for which the 16S identification result was different from the NH card result. These were all unclaimed species by the system. The new NH card met all performance criteria within a 95% confidence interval compared to identification of clinical isolates by 16S.

PubMed Disclaimer

Similar articles

Cited by

References

    1. Alexander, S., and C. Ison. 2005. Evaluation of commercial kits for the identification of Neisseria gonorrhoeae. J. Med. Microbiol. 54827-831. - PubMed
    1. Balganesh, M., M. K. Lalitha, and R. Nathaniel. 2000. Rapid diagnosis of acute pyogenic meningitis by a combined PCR dot-blot assay. Mol. Cell. Probes 1461-69. - PubMed
    1. Barbé, G., M. Babolat, J. M. Boeufgras, D. Monget, and J. Freney. 1994. Evaluation of API NH, a new 2-hour system for identification of Neisseria and Haemophilus species and Moraxella catarrhalis in a routine clinical laboratory. J. Clin. Microbiol. 32187-189. - PMC - PubMed
    1. Benson, D. A., I. Karsch-Mizrachi, D. J. Lipman, J. Ostell, and D. L. Wheeler. 2007. GenBank. Nucleic Acids Res. 35(database issue)D21-D25. - PMC - PubMed
    1. Clarke, S. C. 2006. Detection of Neisseria meningitidis, Streptococcus pneumoniae, and Haemophilus influenzae in blood and cerebrospinal fluid using fluorescence-based PCR. Methods Mol. Biol. 34569-77. - PubMed

MeSH terms

LinkOut - more resources