Mechanical aspects of degree of cement bonding and implant wedge effect
- PMID: 18584929
- DOI: 10.1016/j.clinbiomech.2008.05.003
Mechanical aspects of degree of cement bonding and implant wedge effect
Abstract
Background: The degree of bonding between the femoral stem and cement in total hip replacement remains controversial. Our objective was to determine the wedge effect by debonding and stem taper angle on the structural behavior of axisymmetric stem-cement-bone cylinder models.
Methods: Stainless steel tapered plugs with a rough (i.e. bonded) or smooth (i.e. debonded) surface finish were used to emulate the femoral stem. Three different stem taper angles (5 degrees , 7.5 degrees , 10 degrees ) were used for the debonded constructs. Non-tapered and tapered (7.5 degrees ) aluminum cylindrical shells were used to emulate the diaphyseal and metaphyseal segments of the femur. The cement-aluminum cylinder interface was designed to have a shear strength that simulated bone-cement interfaces ( approximately 8MPa). The test involved applying axial compression at a rate of 0.02mm/s until failure. Six specimens were tested for each combination of the variables. Finite element analysis was used to enhance the understanding of the wedge effect.
Findings: The debonded stems sustained about twice as much load as the bonded stem, regardless of taper angle. The metaphyseal model carried 35-50% greater loads than the diaphyseal models and the stem taper produced significant differences. Based on the finite element analysis, failure was most probably by shear at the cement-bone interface.
Interpretation: Our results in this simplified model suggest that smooth (i.e. debonded) stems have greater failure loads and will incur less slippage or shear failure at the cement-bone interface than rough (i.e. bonded) stems.
Similar articles
-
[Influence of proximal stem geometry and stem-cement interface characteristics on bone and cement stresses in femoral hip arthroplasty: finite element analysis].Rev Chir Orthop Reparatrice Appar Mot. 2003 Apr;89(2):134-43. Rev Chir Orthop Reparatrice Appar Mot. 2003. PMID: 12844057 French.
-
[Mathematical simulation of stem/cement/bone mechanical interactions for Poldi-Cech, CF-30, MS-30 and PFC femoral components].Acta Chir Orthop Traumatol Cech. 2009 Apr;76(2):110-5. Acta Chir Orthop Traumatol Cech. 2009. PMID: 19439130 Czech.
-
[Noncemented total hip arthroplasty: influence of extramedullary parameters on initial implant stability and on bone-implant interface stresses].Rev Chir Orthop Reparatrice Appar Mot. 2000 Oct;86(6):590-7. Rev Chir Orthop Reparatrice Appar Mot. 2000. PMID: 11060433 French.
-
Cement mantle retention: filling the hole.Orthopedics. 2009 Sep;32(9):orthosupersite.com/view.asp?rID=42839. doi: 10.3928/01477447-20090728-16. Orthopedics. 2009. PMID: 19751012 Review.
-
A short stem solution: through small portals.Orthopedics. 2009 Sep;32(9):orthosupersite.com/view.asp?rID=42833. doi: 10.3928/01477447-20090728-09. Orthopedics. 2009. PMID: 19751018 Review.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical