Imaging-therapy computed tomography with quasi-monochromatic X-rays
- PMID: 18586430
- DOI: 10.1016/j.ejrad.2008.04.040
Imaging-therapy computed tomography with quasi-monochromatic X-rays
Abstract
Introduction: Computed tomography (CT) is a widespread and highly precise technique working in the energy range around 50-100 keV. For radiotherapy, however, the MeV energy range enables a better dose distribution. This gap between diagnosis and therapy can be overcome by the use of a modified CT machine in combination with heavy elements targeted to the tumour and used as photoelectric radiation enhancer.
Materials and methods: The experimental setup consists of an X-ray optical module mounted at the exit of the X-ray tube of a clinical CT. The module converts the standard fan-shaped beam into a high intensity, monochromatized and focused beam. The radiation was characterized using an energy-dispersive detection system calibrated by synchrotron radiation and gel dosimetry. The photoelectric radiation enhancement for different elements was calculated and experimentally verified.
Results: The X-ray optical module filters selectively the energy of the tungsten K alpha-emission line (59.3 keV) with a full width at half maximum (FWHM) of 5 keV and focused the radiation onto a focal spot which coincides with the isocentre of the gantry. This results in a steep dose gradient at the centre of rotation qualified for locoregional radiation therapy. The photon energy of the quasi-monochromatic radiation agrees with the energy range of maximal photoelectric dose enhancement for gadolinium and iodine.
Conclusion: An additional X-ray optical module optimized for targeted therapy and photoelectric dose enhancement allows the combination of diagnosis and radiotherapy on a clinical CT.
Similar articles
-
Photoelectric-enhanced radiation therapy with quasi-monochromatic computed tomography.Med Phys. 2009 Jun;36(6):2107-17. doi: 10.1118/1.3125137. Med Phys. 2009. PMID: 19610300
-
Monochromatic x-rays in digital mammography.Invest Radiol. 2005 Jan;40(1):33-9. Invest Radiol. 2005. PMID: 15597018
-
A prototype of a quasi-monochromatic system for mammography applications.Phys Med Biol. 2005 May 21;50(10):2225-40. doi: 10.1088/0031-9155/50/10/003. Epub 2005 Apr 27. Phys Med Biol. 2005. PMID: 15876663
-
Computed tomography with monochromatic x rays.Am J Physiol Imaging. 1992 Jul-Dec;7(3-4):175-93. Am J Physiol Imaging. 1992. PMID: 1343214 Review.
-
[Clinical applications of synchrotron radiation X-ray].Nihon Rinsho. 1994 Sep;52(9):2464-72. Nihon Rinsho. 1994. PMID: 7967098 Review. Japanese.
Cited by
-
Improving x-ray fluorescence signal for benchtop polychromatic cone-beam x-ray fluorescence computed tomography by incident x-ray spectrum optimization: a Monte Carlo study.Med Phys. 2014 Oct;41(10):101906. doi: 10.1118/1.4895004. Med Phys. 2014. PMID: 25281958 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Miscellaneous