Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2008 Jul 21;53(14):3943-69.
doi: 10.1088/0031-9155/53/14/014. Epub 2008 Jun 30.

Efficient array design for sonotherapy

Affiliations

Efficient array design for sonotherapy

Douglas N Stephens et al. Phys Med Biol. .

Abstract

New linear multi-row, multi-frequency arrays have been designed, constructed and tested as fully operational ultrasound probes to produce confocal imaging and therapeutic acoustic intensities with a standard commercial ultrasound imaging system. The triple-array probes and imaging system produce high quality B-mode images with a center row imaging array at 5.3 MHz and sufficient acoustic power with dual therapeutic arrays to produce mild hyperthermia at 1.54 MHz. The therapeutic array pair in the first probe design (termed G3) utilizes a high bandwidth and peak pressure, suitable for mechanical therapies. The second multi-array design (termed G4) has a redesigned therapeutic array pair which is optimized for a high time-averaged power output suitable for mild hyperthermia applications. The 'thermal therapy' design produces more than 4 W of acoustic power from the low-frequency arrays with only a 10.5 degrees C internal rise in temperature after 100 s of continuous use with an unmodified conventional imaging system or substantially longer operation at lower acoustic power. The low-frequency arrays in both probe designs were examined and contrasted for real power transfer efficiency with a KLM model which includes all lossy contributions in the power delivery path from system transmitters to the tissue load. Laboratory verification was successfully performed for the KLM-derived estimates of transducer parallel model acoustic resistance and dissipation resistance, which are the critical design factors for acoustic power output and undesired internal heating, respectively.

PubMed Disclaimer

Figures

Figure 1a
Figure 1a
The multi-array CoLinear common features are two 6mm elevation outside arrays at 1.5MHz and a single 4mm elevation 5.3MHz array (a). The G3 and G4 are different in their outside row designs. The G3 uses a triple layer piezoceramic and two matching layers (b), while the G4 design uses a single layer piezoceramic and single matching layer (c).
Figure 1b
Figure 1b
A representative B-mode image acquired during a sonotherapy treatment of a mouse tumor model as a demonstration of the high level of image quality from the use of the 128 element center imaging array in each of the CoLinear probes. In this particular set up, the Antares system, operating as well in spectral Doppler with the low frequency array pair to produce the mild hyperthermia, is in closed loop control with thermocouple feedback with Doppler-PRF modulation to directly maintain a desired target temperature in the range of 41 to 43 °C.
Figure 2
Figure 2
A generalized system interface for a transducer array. The top panel shows the two components of the real power delivered to the array elements as the desired acoustic (Pa) and undesired dissipative (Pd) power. The lower panel shows the principal loss components in power transmission with a simplified tank circuit model at resonance.
Figure 3
Figure 3
The matrix elements of the KLM model used to characterize the CoLinear designs. A general model for a 2-way entire network is shown in (a), our CoLinear design for 1-way system transmission and tissue loading characteristics in (b), and details of the constituents are shown in (c) as a series of 2X2 matrices.
Figure 4
Figure 4
The KLM model calculated spectral response comparisons for the G3 (1a – 1f) and G4 (2a – 2f) low frequency transducer designs, with computational results at f0 for each case. All paired cases here are performed with unvaried transmit drive voltages (8.5V and 14V, for G3, and G4 respectively) needed to produce 2.1 Watts total acoustic power output with full system and cable. The 1a-2a power spectra with full system and cable pair are derived from (13); the power spectrum pair, 1c-2c, with an ideal voltage source driving the transducers are derived from (19). The efficiency spectra shown in 1b-2b and 1d-2d are derived from (15) and (20) respectively. The “power spectrum efficiency metric” plots of 1e-2e and 1f-2f are derived from (22) and (27) respectively.
Figure 5
Figure 5
The transducer design efficiency (Popt_h) summary spectra for five thicknesses of the graphite composite matching layer in the G4 design, and a single plot comparison for the G3 design with its double layer matching design.
Figure 6
Figure 6
The linear characterization of the two CoLinear designs showing total actual laboratory acoustic power transmitted for a given driving voltage squared.
Figure 7
Figure 7
The acoustic power output spectra, for the G3 design (top) and the G4 design (bottom). The KLM model prediction for power output is stated for f0=1.54MHz, and as a solid line (see also Fig. 4-1a and Fig. 4-2a); the lab results are shown as points with 10% error bars, and with G3 operating as at a conservatively low power level in this test.
Figure 8
Figure 8
The wave shape comparisons of model (dashed) and lab measurements (solid) of B-mode transmissions for the G3 design (top) and the G4 (bottom).
Figure 9
Figure 9
The normalized, unitless correction factor values at the eleven test frequencies used for G3 and G4 estimates of Ra and Rd resistances. The correction for G4 is near unity due to very light loading on the cable and imaging system transmitters.
Figure 10
Figure 10
The acoustic radiation resistance (Ra) and the dissipation resistance (Rd) for the G3 probe design. The solid and dashed traces are the model predictions for Ra and Rd respectively. The circles and crosses are calculated from lab data based on acoustic power measurements and thermistor step responses in long burst transmission tests. The boxed points at 1.54MHz are unique “seed” points that served as a model-derived reference points for absolute scaling of the lab data.
Figure 11
Figure 11
The acoustic radiation electrical resistance (Ra) and the dissipation electrical resistance (Rd) for the G4 probe design. The solid and dashed traces are the model predictions for Ra and Rd respectively. The circles and crosses are calculated from lab data based on acoustic power measurements and thermistor step responses in long burst transmission tests. The boxed points at 1.54MHz are unique pivot points that served as model-derived reference points for absolute scaling of the lab data.
Figure 12
Figure 12
The backing material embedded thermistor temperature response curves for the G3 and G4 designs at various acoustic outputs. Starting at 23 °C the G4 shows only a 10.5 °C rise at 100 seconds of 4W continuous output.
Figure 13
Figure 13
The acoustic impedances Z1 and Z2 which are the loads for their corresponding interfaces for the direction shown.
Figure 14
Figure 14
Comparison for the analytical estimates of equivalent motional resistance for the G3 design model. The KLM model was used to estimate Rpk, the parallel combination of Ra and Rd. The standard analytical estimate for motional resistance, Rpa, is presented along with the modified version, Rpas, and two simplified single expression estimates.
Figure 15
Figure 15
Comparison for the analytical estimates of equivalent motional resistance for the G4 design model. The KLM model was used to estimate Rpk, the parallel combination of Ra and Rd. The standard analytical estimate for motional resistance, Rpa, is presented along with the modified version, Rpas, and two simplified single expression estimates.

References

    1. Allen TM. Liposomes - Opportunities in drug delivery. Drugs. 1997;54:8–14. - PubMed
    1. Chen YC, Wu Sean. Multiple Acoustical Matching Layer Design of Ultrasonic Transducer for Medical Application. Jpn J Appl Phys. 2002;41:6098–107.
    1. Chopra R, Luginbuhl C, Foster FS, Bronskill MJ. Multifrequency ultrasound transducers for conformal interstitial thermal therapy. IEEE Transactions on Ultrasonics Ferroelectrics and Frequency Control. 2003;50:881–9. - PubMed
    1. Cobbold RSC. Foundations of Biomedical Ultrasound. New York, N.Y.: Oxford University Press; 2007.
    1. Desilets CS, Fraser JD, Kino GS. Design of Efficient Broad-Band Piezoelectric Transducers. IEEE Transactions on Sonics and Ultrasonics. 1978;25:115–25.

Publication types