Extravascular lung water volume measurement by a novel lithium-thermal indicator dilution method: comparison of three techniques to post-mortem gravimetry
- PMID: 18607563
- DOI: 10.1007/s00134-008-1207-4
Extravascular lung water volume measurement by a novel lithium-thermal indicator dilution method: comparison of three techniques to post-mortem gravimetry
Abstract
Objective: To compare the lithium-thermal double indicator dilution (Li-thermal), indocyanine green-thermal double indicator dilution (ICG-thermal), single thermal indicator dilution (single-thermal) and gravimetric techniques of extravascular lung water volume (EVLW) measurement in porcine models of acute lung injury.
Design: Two animal models designed to invoke a systemic inflammatory response.
Setting: Laboratory study.
Subjects: A total of 12 immature Deutsches Landschwein pigs.
Interventions: Extravascular lung water volume was measured at four time points using Li-thermal, ICG-thermal and single-thermal techniques. Measurements were performed using existing technology according to manufacturer's instructions. Post-mortem gravimetric EVLW measurements were performed by measuring wet and dry mass of lung tissue. Measurements were compared using the Bland-Altman method. Data are presented as mean (SD).
Measurements and main results: Data were collected in 12 animals and comparison between all 4 techniques was possible in 10 animals. EVLW measured by gravimetry was 9.2 (+/-3.0)ml kg(-1). When compared to gravimetry, both Li-thermal and ICG-thermal techniques showed minimal bias but wide limits of agreement (LOA) [Li-thermal: bias -1.8 ml kg(-1) (LOA +/- 13.1); ICG-thermal bias -1.0 ml kg(-1) (LOA +/- 6.6)]. Comparison between the single-thermal and gravimetric methods identified both considerable bias and wide LOA [+8.5 ml kg(-1) (LOA +/- 14.5)].
Conclusion: Clinically significant differences between EVLW measurements obtained with the gravimetric method and three in vivo indicator dilution techniques were identified. While none of the techniques could be considered ideal, the ICG-thermal method appeared more reliable than either the Li-thermal or single thermal techniques. Further research is required to determine whether the accuracy of the prototype Li-thermal technique can be improved.
Similar articles
-
Comparison of three methods of extravascular lung water volume measurement in patients after cardiac surgery.Crit Care. 2009;13(4):R107. doi: 10.1186/cc7948. Epub 2009 Jul 6. Crit Care. 2009. PMID: 19580649 Free PMC article.
-
Comparison of a single indicator and gravimetric technique for estimation of extravascular lung water in endotoxemic pigs.Crit Care Med. 2006 May;34(5):1437-43. doi: 10.1097/01.CCM.0000215830.48977.29. Crit Care Med. 2006. PMID: 16557161
-
[Comparison of single-indicator thermodilution versus gravimetric measurement in determination of extra-vascular lung water in dogs with acute respiratory distress syndrome].Zhongguo Wei Zhong Bing Ji Jiu Yi Xue. 2006 Jun;18(6):327-30. Zhongguo Wei Zhong Bing Ji Jiu Yi Xue. 2006. PMID: 16784555 Chinese.
-
Bedside assessment of extravascular lung water by dilution methods: temptations and pitfalls.Crit Care Med. 2007 Apr;35(4):1186-92. doi: 10.1097/01.CCM.0000259539.49339.66. Crit Care Med. 2007. PMID: 17334247 Review.
-
Measurement of extravascular lung water using the single indicator method in patients: research and potential clinical value.Am J Physiol Lung Cell Mol Physiol. 2009 Oct;297(4):L547-58. doi: 10.1152/ajplung.00127.2009. Epub 2009 Jul 17. Am J Physiol Lung Cell Mol Physiol. 2009. PMID: 19617309 Free PMC article. Review.
Cited by
-
Transthoracic ultrasound assessment of B-lines for identifying the increment of extravascular lung water in shock patients requiring fluid resuscitation.Indian J Crit Care Med. 2014 Apr;18(4):195-9. doi: 10.4103/0972-5229.130569. Indian J Crit Care Med. 2014. PMID: 24872647 Free PMC article.
-
Lung ultrasound is a reliable method for evaluating extravascular lung water volume in rodents.BMC Anesthesiol. 2015 Nov 12;15:162. doi: 10.1186/s12871-015-0146-1. BMC Anesthesiol. 2015. PMID: 26563098 Free PMC article.
-
Comparison of three methods of extravascular lung water volume measurement in patients after cardiac surgery.Crit Care. 2009;13(4):R107. doi: 10.1186/cc7948. Epub 2009 Jul 6. Crit Care. 2009. PMID: 19580649 Free PMC article.
-
Extravascular lung water index measurement in critically ill children does not correlate with a chest x-ray score of pulmonary edema.Crit Care. 2010;14(3):R105. doi: 10.1186/cc9054. Epub 2010 Jun 8. Crit Care. 2010. PMID: 20529308 Free PMC article.
-
Estimation of extravascular lung water using the transpulmonary ultrasound dilution (TPUD) method: a validation study in neonatal lambs.J Clin Monit Comput. 2016 Dec;30(6):985-994. doi: 10.1007/s10877-015-9803-7. Epub 2015 Nov 12. J Clin Monit Comput. 2016. PMID: 26563187 Free PMC article.
References
Publication types
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources