A uniaxial bioMEMS device for imaging single cell response during quantitative force-displacement measurements
- PMID: 18648937
- DOI: 10.1007/s10544-008-9202-7
A uniaxial bioMEMS device for imaging single cell response during quantitative force-displacement measurements
Abstract
A microfabricated device has been developed for imaging of a single, adherent cell while quantifying force under an applied displacement. The device works in a fashion similar to that of a displacement-controlled uniaxial tensile machine. The device was calibrated using a tipless atomic force microscope (AFM) cantilever and shows excellent agreement with the calculated spring constant. A step input was applied to a single, adherent fibroblast cell and the viscoelastic response was characterized with a mechanical model. The adherent fibroblast was imaged by use of epifluorescence and phase contrast techniques.
MeSH terms
LinkOut - more resources
Full Text Sources