An evaluation of microbiologic contamination on phosphor plates in a dental school
Abstract
Objective: The objective of this study was to determine if phosphor plates used in predoctoral clinics are microbiologically contaminated and to identify the source of contamination.
Study design: Forty-five of 300 phosphor plates (15%) were randomly selected for examination. The plates were pressed into individual blood agar plates, were incubated using standard techniques at 37 degrees C, and were monitored for 72 hours. The number, size, distribution, and variety of resulting colonies were noted. A representative of each type of colony was selected to be Gram stained using the standard technique.
Results: Of the plates, 42.2% were uncontaminated, 57.8% yielded bacterial colonies, and 15.6% of those colonies demonstrated hemolytic growth. The hemolytic growth included combined alpha and beta hemolysis and beta only hemolysis. Six colonies were gram-positive rods and 7 were gram-positive cocci.
Conclusion: Meticulous infection-control techniques are inevitable and continuous reinforcement and training for staff and students are mandatory. Periodic gas sterilization of phosphor plates may be necessary.
Similar articles
-
Microbial contamination in intraoral phosphor storage plates: the dilemma.Clin Oral Investig. 2017 Jan;21(1):301-307. doi: 10.1007/s00784-016-1790-7. Epub 2016 Mar 16. Clin Oral Investig. 2017. PMID: 26984823
-
An evaluation of microbiologic contamination on a phosphor plate system: is weekly gas sterilization enough?Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2010 Mar;109(3):457-62. doi: 10.1016/j.tripleo.2009.09.035. Epub 2010 Jan 8. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2010. PMID: 20060337
-
Infection control in digital intraoral radiography: evaluation of microbiological contamination of photostimulable phosphor plates in barrier envelopes.J Can Dent Assoc. 2011;77:b93. J Can Dent Assoc. 2011. PMID: 21929948
-
Do contaminated dental unit waterlines pose a risk of infection?J Dent. 2007 Sep;35(9):712-20. doi: 10.1016/j.jdent.2007.06.002. Epub 2007 Aug 6. J Dent. 2007. PMID: 17689168 Review.
-
Exposure indicator degradation from CR plate processing delays.Radiol Technol. 2009 May-Jun;80(5):401-9. Radiol Technol. 2009. PMID: 19457844 Review.
Cited by
-
Effect of enveloping and disinfection methods on artefact formation on enveloped PSP plate images.Oral Radiol. 2022 Oct;38(4):558-564. doi: 10.1007/s11282-022-00587-1. Epub 2022 Jan 22. Oral Radiol. 2022. PMID: 35064886 Free PMC article.
-
Microbial contamination in intraoral phosphor storage plates: the dilemma.Clin Oral Investig. 2017 Jan;21(1):301-307. doi: 10.1007/s00784-016-1790-7. Epub 2016 Mar 16. Clin Oral Investig. 2017. PMID: 26984823
-
Perforation rate of intraoral barriers for direct digital radiography.Dentomaxillofac Radiol. 2015;44(3):20140245. doi: 10.1259/dmfr.20140245. Epub 2014 Nov 20. Dentomaxillofac Radiol. 2015. PMID: 25410788 Free PMC article.
-
Infection control and patient discomfort with an alternative plastic barrier in intraoral digital radiography.Dentomaxillofac Radiol. 2017 Feb;46(2):20160253. doi: 10.1259/dmfr.20160253. Epub 2016 Dec 20. Dentomaxillofac Radiol. 2017. PMID: 27996305 Free PMC article.
-
The effectiveness of a pre-procedural mouthrinse in reducing bacteria on radiographic phosphor plates.Imaging Sci Dent. 2014 Jun;44(2):149-54. doi: 10.5624/isd.2014.44.2.149. Epub 2014 Jun 11. Imaging Sci Dent. 2014. PMID: 24944965 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources