Abstract
Aim: Matrix perimetry uses frequency-doubling technology (FDT) incorporated into a 5 degrees test target. This permits testing of the same number of locations within a defined visual field as standard automated perimetry (SAP) and may improve performance compared with original FDT perimetry. This study investigates the performance of Humphrey Matrix perimetry for detecting glaucomatous visual-field loss.
Design: Prospective case control study.
Methods: We recruited 115 participants with glaucomatous visual-field loss and 33 normal controls from an urban glaucoma practice. Each participant performed SITA 24-2 SAP then threshold 24-2 Matrix perimetry. Severity of visual-field loss was defined using SAP mean deviation (MD) as early (MD >-6 dB), moderate (MD -6 to -12 dB) or advanced (MD <-12 dB). The sensitivity and specificity of Humphrey Matrix perimetry were calculated for different automated indices.
Results: The matrix perimetry sensitivity and specificity were up to 100% for moderate and advanced glaucomatous visual-field loss. A receiver operator characteristic area under the curve (AUC) analysis revealed MD to be slightly better than pattern standard deviation (PSD) for defining moderate (AUC: MD 0.997; PSD 0.987) and advanced defects (AUC: MD 1.000; PSD 0.987). Matrix was less sensitive (up to 87.3%) for detecting early glaucomatous visual-field loss compared with SITA 24-2 SAP (AUC: PSD 0.948; MD 0.910).
Conclusions: Matrix perimetry is excellent for detection of moderate to advanced glaucomatous visual-field loss but may miss some early defects. It may be well suited to following progression of early to moderate field loss because of a smaller target size compared with original FDT perimetry.
Similar articles
-
Frequency-doubling perimetry: comparison with standard automated perimetry to detect glaucoma.Am J Ophthalmol. 2007 Feb;143(2):263-271. doi: 10.1016/j.ajo.2006.10.033. Epub 2006 Nov 27. Am J Ophthalmol. 2007. PMID: 17178091 Clinical Trial.
-
Threshold and variability properties of matrix frequency-doubling technology and standard automated perimetry in glaucoma.Invest Ophthalmol Vis Sci. 2005 Jul;46(7):2451-7. doi: 10.1167/iovs.05-0135. Invest Ophthalmol Vis Sci. 2005. PMID: 15980235
-
Detectability of glaucomatous changes using SAP, FDT, flicker perimetry, and OCT.J Glaucoma. 2009 Feb;18(2):165-71. doi: 10.1097/IJG.0b013e318179f7ca. J Glaucoma. 2009. PMID: 19225357
-
Effect of cataract extraction on frequency doubling technology perimetry in patients with glaucoma.Br J Ophthalmol. 2005 Dec;89(12):1569-71. doi: 10.1136/bjo.2005.080655. Br J Ophthalmol. 2005. PMID: 16299131 Free PMC article. Review.
-
Short-wavelength automated perimetry and frequency-doubling technology perimetry in glaucoma.Prog Brain Res. 2008;173:101-24. doi: 10.1016/S0079-6123(08)01108-4. Prog Brain Res. 2008. PMID: 18929104 Review.
Cited by
-
Glaucoma diagnostic performance of humphrey matrix and standard automated perimetry.Jpn J Ophthalmol. 2009 Sep;53(5):482-5. doi: 10.1007/s10384-009-0717-0. Epub 2009 Oct 22. Jpn J Ophthalmol. 2009. PMID: 19847602
-
Standard automated perimetry versus matrix frequency doubling technology perimetry in subjects with ocular hypertension and healthy control subjects.PLoS One. 2013;8(2):e57663. doi: 10.1371/journal.pone.0057663. Epub 2013 Feb 28. PLoS One. 2013. PMID: 23469043 Free PMC article.
-
Predicting progression of glaucoma from rates of frequency doubling technology perimetry change.Ophthalmology. 2014 Feb;121(2):498-507. doi: 10.1016/j.ophtha.2013.09.016. Epub 2013 Nov 26. Ophthalmology. 2014. PMID: 24289917 Free PMC article.
-
Development of an LCD-Based Visual Field System.J Korean Med Sci. 2018 Jan 15;33(3):e19. doi: 10.3346/jkms.2018.33.e19. J Korean Med Sci. 2018. PMID: 29215806 Free PMC article. Clinical Trial.
-
Responses of primate retinal ganglion cells to perimetric stimuli.Invest Ophthalmol Vis Sci. 2011 Feb 9;52(2):764-71. doi: 10.1167/iovs.10-6158. Invest Ophthalmol Vis Sci. 2011. PMID: 20881286 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Miscellaneous