Humphrey matrix frequency doubling perimetry for detection of visual-field defects in open-angle glaucoma

Affiliations

Humphrey matrix frequency doubling perimetry for detection of visual-field defects in open-angle glaucoma

C I Clement et al. Br J Ophthalmol. 2009 May.

Abstract

Aim: Matrix perimetry uses frequency-doubling technology (FDT) incorporated into a 5 degrees test target. This permits testing of the same number of locations within a defined visual field as standard automated perimetry (SAP) and may improve performance compared with original FDT perimetry. This study investigates the performance of Humphrey Matrix perimetry for detecting glaucomatous visual-field loss.

Design: Prospective case control study.

Methods: We recruited 115 participants with glaucomatous visual-field loss and 33 normal controls from an urban glaucoma practice. Each participant performed SITA 24-2 SAP then threshold 24-2 Matrix perimetry. Severity of visual-field loss was defined using SAP mean deviation (MD) as early (MD >-6 dB), moderate (MD -6 to -12 dB) or advanced (MD <-12 dB). The sensitivity and specificity of Humphrey Matrix perimetry were calculated for different automated indices.

Results: The matrix perimetry sensitivity and specificity were up to 100% for moderate and advanced glaucomatous visual-field loss. A receiver operator characteristic area under the curve (AUC) analysis revealed MD to be slightly better than pattern standard deviation (PSD) for defining moderate (AUC: MD 0.997; PSD 0.987) and advanced defects (AUC: MD 1.000; PSD 0.987). Matrix was less sensitive (up to 87.3%) for detecting early glaucomatous visual-field loss compared with SITA 24-2 SAP (AUC: PSD 0.948; MD 0.910).

Conclusions: Matrix perimetry is excellent for detection of moderate to advanced glaucomatous visual-field loss but may miss some early defects. It may be well suited to following progression of early to moderate field loss because of a smaller target size compared with original FDT perimetry.