Implementation of DART and DESI ionization on a fieldable mass spectrometer
- PMID: 18703351
- DOI: 10.1016/j.jasms.2008.06.028
Implementation of DART and DESI ionization on a fieldable mass spectrometer
Abstract
A recently developed prototype mobile laboratory mass spectrometer, incorporating an atmospheric pressure ionization (API) interface, is described. This system takes advantage of the small size, lower voltage requirements, and tandem MS abilities of the cylindrical ion trap mass analyzer. The prototype API MS uses small, low-power pumps to fit into a 0.1-m(3) self-contained package weighing <45 kg. This instrument has been adapted to allow rapid interfacing to electrospray ionization, desorption electrospray ionization, and direct analysis in real-time sources. Initial data indicate that these techniques provide rapid detection and identification of compounds for quality control, homeland security, and forensic applications. In addition, this instrument is self-contained and compact, making it ideally extensible to mobile laboratory and field analyses. Initial MS and MS/MS data for analyses of drugs, food, and explosives are presented herein.
Similar articles
-
Desorption electrospray ionization of explosives on surfaces: sensitivity and selectivity enhancement by reactive desorption electrospray ionization.Anal Chem. 2005 Nov 1;77(21):6755-64. doi: 10.1021/ac050995+. Anal Chem. 2005. PMID: 16255571
-
Non-proximate detection of small and large molecules by desorption electrospray ionization and desorption atmospheric pressure chemical ionization mass spectrometry: instrumentation and applications in forensics, chemistry, and biology.Anal Chem. 2007 Sep 15;79(18):7069-77. doi: 10.1021/ac0707939. Epub 2007 Aug 14. Anal Chem. 2007. PMID: 17696318
-
Atmospheric pressure ionization in a miniature mass spectrometer.Anal Chem. 2005 May 1;77(9):2928-39. doi: 10.1021/ac0481708. Anal Chem. 2005. PMID: 15859613
-
Surface analysis of lipids by mass spectrometry: more than just imaging.Prog Lipid Res. 2013 Oct;52(4):329-53. doi: 10.1016/j.plipres.2013.04.005. Epub 2013 Apr 24. Prog Lipid Res. 2013. PMID: 23623802 Review.
-
FIELDABLE MASS SPECTROMETRY FOR FORENSIC SCIENCE, HOMELAND SECURITY, AND DEFENSE APPLICATIONS.Mass Spectrom Rev. 2021 Sep;40(5):628-646. doi: 10.1002/mas.21646. Epub 2020 Jul 28. Mass Spectrom Rev. 2021. PMID: 32722885 Review.
Cited by
-
A Theoretical Method for Characterizing Nonlinear Effects in Paul Traps with Added Octopole Field.J Am Soc Mass Spectrom. 2015 Aug;26(8):1338-48. doi: 10.1007/s13361-015-1145-7. Epub 2015 Apr 30. J Am Soc Mass Spectrom. 2015. PMID: 25924875
-
Proof of Concept Coded Aperture Miniature Mass Spectrometer Using a Cycloidal Sector Mass Analyzer, a Carbon Nanotube (CNT) Field Emission Electron Ionization Source, and an Array Detector.J Am Soc Mass Spectrom. 2018 Feb;29(2):360-372. doi: 10.1007/s13361-017-1820-y. Epub 2017 Oct 19. J Am Soc Mass Spectrom. 2018. PMID: 29052038
-
What can we learn from ambient ionization techniques?J Am Soc Mass Spectrom. 2009 Nov;20(11):1947-63. doi: 10.1016/j.jasms.2009.07.025. Epub 2009 Aug 13. J Am Soc Mass Spectrom. 2009. PMID: 19748284 Review.
-
Matrix-Assisted Ionization on a Portable Mass Spectrometer: Analysis Directly from Biological and Synthetic Materials.Anal Chem. 2016 Nov 15;88(22):10831-10836. doi: 10.1021/acs.analchem.6b00304. Epub 2016 Nov 3. Anal Chem. 2016. PMID: 26938428 Free PMC article.
-
Recent advances in ambient mass spectrometry of trace explosives.Analyst. 2018 Apr 30;143(9):1948-1969. doi: 10.1039/c7an02066j. Analyst. 2018. PMID: 29664070 Free PMC article. Review.
References
LinkOut - more resources
Full Text Sources
Other Literature Sources