Effect of central corneal thickness and corneal hysteresis on tonometry as measured by dynamic contour tonometry, ocular response analyzer, and Goldmann tonometry in glaucomatous eyes
- PMID: 18703945
- DOI: 10.1097/IJG.0b013e31815c3ad3
Effect of central corneal thickness and corneal hysteresis on tonometry as measured by dynamic contour tonometry, ocular response analyzer, and Goldmann tonometry in glaucomatous eyes
Erratum in
- J Glaucoma. 2008 Oct-Nov;17(7):604. Annette, Hager [corrected to Hager, Annette]; Kristina, Loge [corrected to Loge, Kristina]; Bernd, Schroeder [corrected to Schroeder, Bernd]; Mark-Oliver, Füllhas [corrected to Füllhas, Mark-Oliver]; Wolfgang, Wiegand [corrected to Wiegand, Wolfgang]
Abstract
Purpose: The dynamic contour tonometer (DCT; Pascal tonometer) and the Ocular Response Analyzer (ORA) are novel tonometers designed to measure intraocular pressure (IOP) independent of corneal properties and central corneal thickness (CCT), respectively. We wanted to compare the corneal compensated IOP (IOPcc) as measured by ORA with IOP values measured by DCT and Goldmann applanation tonometry (GAT) with respect to IOP readings and the influence of corneal hysteresis (CH) and CCT in glaucoma patients.
Patients and methods: In a study of 94 glaucomatous eyes, IOP measurements by ORA, DCT, and GAT were compared, and the effects of CCT and CH were analyzed. All measurements were taken by 1 of the authors only.
Results: The Mean CCT was 550+/-44 mum and mean CH 10.24+/-3.3 mm Hg. The mean value for IOPcc was 17.94+/-5.9 mm Hg, 15.14+/-3.7 mm Hg for DCT, and 14.3+/-4.3 mm Hg for GAT. The mean difference was 3.65+/-3.85 mm Hg between IOPcc and GAT and 2.80+/-4.9 mm Hg between IOPcc and DCT (P<0.001) and -0.85+/-3.3 mm Hg between GAT and DCT (P=0.015). None of the methods was related to CCT. In contrast, CH was highly statistically significant related to IOPcc (P<0.0001), whereas GAT and DCT showed only weak relation to CH (P=0.05).
Conclusions: IOP measurements with the ORA are significantly higher than DCT or GAT with no apparent linear correction factor. To date, we cannot estimate the impact of differences in CH in relation to IOP and its measurement.
Similar articles
-
The relationship between diurnal variations in intraocular pressure measurements and central corneal thickness and corneal hysteresis.Invest Ophthalmol Vis Sci. 2009 Sep;50(9):4229-36. doi: 10.1167/iovs.08-2955. Epub 2009 Apr 30. Invest Ophthalmol Vis Sci. 2009. PMID: 19407025
-
Intraocular pressure measurement precision with the Goldmann applanation, dynamic contour, and ocular response analyzer tonometers.Ophthalmology. 2010 Apr;117(4):730-7. doi: 10.1016/j.ophtha.2009.09.020. Epub 2010 Feb 1. Ophthalmology. 2010. PMID: 20122737 Clinical Trial.
-
Corneal thickness- and age-related biomechanical properties of the cornea measured with the ocular response analyzer.Invest Ophthalmol Vis Sci. 2006 Dec;47(12):5337-47. doi: 10.1167/iovs.06-0557. Invest Ophthalmol Vis Sci. 2006. PMID: 17122122
-
[What to do if the intraocular pressure measurement does not appear reliable].J Fr Ophtalmol. 2010 Apr;33(4):279-84. doi: 10.1016/j.jfo.2010.02.007. Epub 2010 Mar 29. J Fr Ophtalmol. 2010. PMID: 20347507 Review. French.
-
New ways to measure intraocular pressure.Curr Opin Ophthalmol. 2008 Mar;19(2):122-6. doi: 10.1097/ICU.0b013e3282f391ae. Curr Opin Ophthalmol. 2008. PMID: 18301285 Review.
Cited by
-
Detecting IOP Fluctuations in Glaucoma Patients.Open Ophthalmol J. 2016 Feb 29;10:44-55. doi: 10.2174/1874364101610010044. eCollection 2016. Open Ophthalmol J. 2016. PMID: 27014387 Free PMC article.
-
Distribution of central corneal thickness and its association with ocular parameters in a large central European cohort: the Gutenberg health study.PLoS One. 2013 Aug 1;8(8):e66158. doi: 10.1371/journal.pone.0066158. Print 2013. PLoS One. 2013. PMID: 23936291 Free PMC article.
-
Repeatability of the Novel Intraocular Pressure Measurement From Corvis ST.Transl Vis Sci Technol. 2019 Jun 24;8(3):48. doi: 10.1167/tvst.8.3.48. eCollection 2019 May. Transl Vis Sci Technol. 2019. PMID: 31275733 Free PMC article.
-
Tonometers-which one should I use?Eye (Lond). 2018 May;32(5):931-937. doi: 10.1038/s41433-018-0040-4. Epub 2018 Feb 19. Eye (Lond). 2018. PMID: 29456251 Free PMC article. Review.
-
Comparison of Goldmann and Pascal tonometry in relation to corneal hysteresis and central corneal thickness in nonglaucomatous eyes.Clin Ophthalmol. 2011;5:1071-7. doi: 10.2147/OPTH.S23086. Epub 2011 Aug 3. Clin Ophthalmol. 2011. PMID: 21847339 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical