The Netherlands protocol for standardisation and quantification of FDG whole body PET studies in multi-centre trials
- PMID: 18704407
- DOI: 10.1007/s00259-008-0874-2
The Netherlands protocol for standardisation and quantification of FDG whole body PET studies in multi-centre trials
Erratum in
- Eur J Nucl Med Mol Imaging. 2008 Dec;35(12):2342
Abstract
Introduction: Several studies have shown the usefulness of positron emission tomography (PET) quantification using standardised uptake values (SUV) for diagnosis and staging, prognosis and response monitoring. Many factors affect SUV, such as patient preparation procedures, scan acquisition, image reconstruction and data analysis settings, and the variability in methodology across centres prohibits exchange of SUV data. Therefore, standardisation of 2-[(18)F] fluoro-2-deoxy-D-glucose (FDG) PET whole body procedures is required in multi-centre trials.
Methods: A protocol for standardisation of quantitative FDG whole body PET studies in the Netherlands (NL) was defined. This protocol is based on standardisation of: (1) patient preparation; (2) matching of scan statistics by prescribing dosage as function of patient weight, scan time per bed position, percentage of bed overlap and image acquisition mode (2D or 3D); (3) matching of image resolution by prescribing reconstruction settings for each type of scanner; (4) matching of data analysis procedure by defining volume of interest methods and SUV calculations and; (5) finally, a multi-centre QC procedure is defined using a 20-cm diameter phantom for verification of scanner calibration and the NEMA NU 2 2001 Image Quality phantom for verification of activity concentration recoveries (i.e., verification of image resolution and reconstruction convergence).
Discussion: This paper describes a protocol for standardization of quantitative FDG whole body multi-centre PET studies.
Conclusion: The protocol was successfully implemented in the Netherlands and has been approved by the Netherlands Society of Nuclear Medicine.
Similar articles
-
Quantification of FDG PET studies using standardised uptake values in multi-centre trials: effects of image reconstruction, resolution and ROI definition parameters.Eur J Nucl Med Mol Imaging. 2007 Mar;34(3):392-404. doi: 10.1007/s00259-006-0224-1. Epub 2006 Oct 11. Eur J Nucl Med Mol Imaging. 2007. PMID: 17033848
-
Standardization of image quality across multiple centers by optimization of acquisition and reconstruction parameters with interim FDG-PET/CT for evaluating diffuse large B cell lymphoma.Ann Nucl Med. 2013 Apr;27(3):225-32. doi: 10.1007/s12149-012-0676-2. Epub 2012 Dec 22. Ann Nucl Med. 2013. PMID: 23264065
-
Interim positron emission tomography scan in multi-center studies: optimization of visual and quantitative assessments.Leuk Lymphoma. 2009 Nov;50(11):1748-9. doi: 10.3109/10428190903308049. Leuk Lymphoma. 2009. PMID: 19863171
-
Variability and uncertainty of 18F-FDG PET imaging protocols for assessing inflammation in atherosclerosis: suggestions for improvement.J Nucl Med. 2015 Apr;56(4):552-9. doi: 10.2967/jnumed.114.142596. Epub 2015 Feb 26. J Nucl Med. 2015. PMID: 25722452 Review.
-
Standardization and quantification in FDG-PET/CT imaging for staging and restaging of malignant disease.PET Clin. 2014 Apr;9(2):117-27. doi: 10.1016/j.cpet.2013.10.003. Epub 2013 Dec 5. PET Clin. 2014. PMID: 25030277 Review.
Cited by
-
Harmonizing SUVs in multicentre trials when using different generation PET systems: prospective validation in non-small cell lung cancer patients.Eur J Nucl Med Mol Imaging. 2013 Jul;40(7):985-96. doi: 10.1007/s00259-013-2391-1. Epub 2013 Apr 6. Eur J Nucl Med Mol Imaging. 2013. PMID: 23564036 Free PMC article. Clinical Trial.
-
Harmonization of standard uptake values across different positron emission tomography/computed tomography systems and different reconstruction algorithms: validation in oncology patients.EJNMMI Phys. 2023 Mar 15;10(1):19. doi: 10.1186/s40658-023-00540-z. EJNMMI Phys. 2023. PMID: 36920590 Free PMC article.
-
Relationship between semiquantitative 18F-fluorodeoxyglucose positron emission tomography metrics and necrosis in classical Hodgkin lymphoma.Sci Rep. 2019 Jul 30;9(1):11073. doi: 10.1038/s41598-019-47453-5. Sci Rep. 2019. PMID: 31363153 Free PMC article.
-
Methodological considerations in quantification of oncological FDG PET studies.Eur J Nucl Med Mol Imaging. 2010 Jul;37(7):1408-25. doi: 10.1007/s00259-009-1306-7. Epub 2009 Nov 20. Eur J Nucl Med Mol Imaging. 2010. PMID: 19936745 Free PMC article. Review.
-
The use of FDG-PET to target tumors by radiotherapy.Strahlenther Onkol. 2010 Sep;186(9):471-81. doi: 10.1007/s00066-010-2150-1. Epub 2010 Aug 30. Strahlenther Onkol. 2010. PMID: 20814658 Review.
References
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Research Materials