. 2008 Dec;35(12):2320-33.
doi: 10.1007/s00259-008-0874-2. Epub 2008 Aug 15.

The Netherlands protocol for standardisation and quantification of FDG whole body PET studies in multi-centre trials

Affiliations

The Netherlands protocol for standardisation and quantification of FDG whole body PET studies in multi-centre trials

Ronald Boellaard et al. Eur J Nucl Med Mol Imaging. 2008 Dec.

Erratum in

  • Eur J Nucl Med Mol Imaging. 2008 Dec;35(12):2342

Abstract

Introduction: Several studies have shown the usefulness of positron emission tomography (PET) quantification using standardised uptake values (SUV) for diagnosis and staging, prognosis and response monitoring. Many factors affect SUV, such as patient preparation procedures, scan acquisition, image reconstruction and data analysis settings, and the variability in methodology across centres prohibits exchange of SUV data. Therefore, standardisation of 2-[(18)F] fluoro-2-deoxy-D-glucose (FDG) PET whole body procedures is required in multi-centre trials.

Methods: A protocol for standardisation of quantitative FDG whole body PET studies in the Netherlands (NL) was defined. This protocol is based on standardisation of: (1) patient preparation; (2) matching of scan statistics by prescribing dosage as function of patient weight, scan time per bed position, percentage of bed overlap and image acquisition mode (2D or 3D); (3) matching of image resolution by prescribing reconstruction settings for each type of scanner; (4) matching of data analysis procedure by defining volume of interest methods and SUV calculations and; (5) finally, a multi-centre QC procedure is defined using a 20-cm diameter phantom for verification of scanner calibration and the NEMA NU 2 2001 Image Quality phantom for verification of activity concentration recoveries (i.e., verification of image resolution and reconstruction convergence).

Discussion: This paper describes a protocol for standardization of quantitative FDG whole body multi-centre PET studies.

Conclusion: The protocol was successfully implemented in the Netherlands and has been approved by the Netherlands Society of Nuclear Medicine.

PubMed Disclaimer

Similar articles

Cited by

References

    1. J Nucl Med. 2002 Oct;43(10):1339-42 - PubMed
    1. J Nucl Med Technol. 2003 Dec;31(4):206-9 - PubMed
    1. AJR Am J Roentgenol. 2002 Aug;179(2):477-81 - PubMed
    1. Nucl Med Commun. 2007 Jun;28(6):485-93 - PubMed
    1. J Nucl Med. 2005 Jun;46(6):983-95 - PubMed

Substances

LinkOut - more resources