Tibial post wear in posterior-stabilized knee replacements is design-dependent
- PMID: 18704615
- PMCID: PMC2565057
- DOI: 10.1007/s11999-008-0422-1
Tibial post wear in posterior-stabilized knee replacements is design-dependent
Abstract
Polyethylene tibial post wear in posterior-stabilized knee designs is a major problem. The Insall-Burstein II (IB PS II) reportedly has severe anterior wear of the post in retrieved implants. We hypothesized the more anterior placement in the IB PS II would be reflected in greater wear at the anterior face than the IB PS I. We examined 234 retrieved inserts using subjective scales to grade post damage and wear. Of the IB PS II inserts, 38% demonstrated severe wear compared with only 25% of IB PS I inserts. The most prevalent damage location for the IB PS II was the anterior face, whereas the IB PS I sustained wear mainly on the medial face. While the IB PS post was not designed to constrain posterior femoral displacement, our observations confirm contact in hyperextension or other paradoxic anterior tibial translation is common and design-dependent. Minimizing wear and damage through proper post placement and changes in implant design to anticipate contact on the anterior post should be considered for future posterior stabilized knee replacements. These changes cannot occur in isolation, however, because changes in post placement and design also depend on their relation to the shape and location of the tibial bearing surfaces.
Figures



Similar articles
-
Polished trays reduce backside wear independent of post location in posterior-stabilized TKAs.Clin Orthop Relat Res. 2014 Aug;472(8):2477-82. doi: 10.1007/s11999-014-3621-y. Epub 2014 Apr 15. Clin Orthop Relat Res. 2014. PMID: 24733449 Free PMC article.
-
Implant design influences tibial post wear damage in posterior-stabilized knees.Clin Orthop Relat Res. 2011 Jan;469(1):160-7. doi: 10.1007/s11999-010-1515-1. Clin Orthop Relat Res. 2011. PMID: 20711697 Free PMC article.
-
Factors Affecting Wear of Constrained Polyethylene Tibial Inserts in Total Knee Arthroplasty.J Arthroplasty. 2016 Jun;31(6):1340-1345. doi: 10.1016/j.arth.2015.12.011. Epub 2015 Dec 17. J Arthroplasty. 2016. PMID: 26777549
-
Cemented posterior stabilized total knee arthroplasty.J Arthroplasty. 2004 Jun;19(4 Suppl 1):17-21. doi: 10.1016/j.arth.2004.02.014. J Arthroplasty. 2004. PMID: 15190543 Review.
-
Posterior-stabilized inserts are preferable to cruciate-substituting ultracongruent inserts due to more favourable kinematics and stability.Knee Surg Sports Traumatol Arthrosc. 2018 Nov;26(11):3300-3310. doi: 10.1007/s00167-018-4872-z. Epub 2018 Feb 19. Knee Surg Sports Traumatol Arthrosc. 2018. PMID: 29459998
Cited by
-
Surface damage versus tibial polyethylene insert conformity: a retrieval study.Clin Orthop Relat Res. 2012 Jul;470(7):1814-25. doi: 10.1007/s11999-012-2274-y. Clin Orthop Relat Res. 2012. PMID: 22395868 Free PMC article.
-
Polished trays reduce backside wear independent of post location in posterior-stabilized TKAs.Clin Orthop Relat Res. 2014 Aug;472(8):2477-82. doi: 10.1007/s11999-014-3621-y. Epub 2014 Apr 15. Clin Orthop Relat Res. 2014. PMID: 24733449 Free PMC article.
-
Design Modifications of the Posterior-Stabilized Knee System May Reduce Anterior Knee Pain and Complications Following Total Knee Replacement.HSS J. 2020 Dec;16(Suppl 2):344-348. doi: 10.1007/s11420-019-09742-7. Epub 2020 Jan 18. HSS J. 2020. PMID: 33380967 Free PMC article.
-
Rotating-platform has no surface damage advantage over fixed-bearing TKA.Clin Orthop Relat Res. 2013 Jan;471(1):76-85. doi: 10.1007/s11999-012-2530-1. Clin Orthop Relat Res. 2013. PMID: 22948523 Free PMC article.
-
Nontraumatic fracture of the polyethylene tibial post in a bi-cruciate stabilized total knee prosthesis.Arthroplast Today. 2019 Jul 23;5(3):284-287. doi: 10.1016/j.artd.2019.06.001. eCollection 2019 Sep. Arthroplast Today. 2019. PMID: 31516966 Free PMC article.
References
-
- {'text': '', 'ref_index': 1, 'ids': [{'type': 'PubMed', 'value': '12479337', 'is_inner': True, 'url': 'https://pubmed.ncbi.nlm.nih.gov/12479337/'}]}
- Banks SA, Harman MK, Hodge WA. Mechanism of anterior impingement damage in total knee arthroplasty. J Bone Joint Surg Am. 2002;84(Suppl 2):37–42. - PubMed
-
- {'text': '', 'ref_index': 1, 'ids': [{'type': 'DOI', 'value': '10.1016/j.arth.2004.04.011', 'is_inner': False, 'url': 'https://doi.org/10.1016/j.arth.2004.04.011'}, {'type': 'PubMed', 'value': '15483794', 'is_inner': True, 'url': 'https://pubmed.ncbi.nlm.nih.gov/15483794/'}]}
- Banks SA, Hodge WA. Design and activity dependence of kinematics in fixed and mobile-bearing knee arthroplasties. J Arthroplasty. 2004;19:809–816. - PubMed
-
- {'text': '', 'ref_index': 1, 'ids': [{'type': 'DOI', 'value': '10.1016/j.arth.2004.02.026', 'is_inner': False, 'url': 'https://doi.org/10.1016/j.arth.2004.02.026'}, {'type': 'PubMed', 'value': '15284990', 'is_inner': True, 'url': 'https://pubmed.ncbi.nlm.nih.gov/15284990/'}]}
- Clarke HD, Math KR, Scuderi GR. Polyethylene post failure in posterior stabilized total knee arthroplasty. J Arthroplasty. 2004;19:652–657. - PubMed
-
- {'text': '', 'ref_index': 1, 'ids': [{'type': 'PubMed', 'value': '9111404', 'is_inner': True, 'url': 'https://pubmed.ncbi.nlm.nih.gov/9111404/'}]}
- Diduch DR, Insall JN, Scott WN, Scuderi GR, Font-Rodriguez D. Total knee replacement in young, active patients. Long-term follow-up and functional outcome. J Bone Joint Surg Am. 1997;79:575–582. - PubMed
-
- {'text': '', 'ref_index': 1, 'ids': [{'type': 'PubMed', 'value': '2805502', 'is_inner': True, 'url': 'https://pubmed.ncbi.nlm.nih.gov/2805502/'}]}