Comparison of NaI(Tl), CdTe, and HgI2 surgical probes: physical characterization
- PMID: 1870478
- DOI: 10.1118/1.596717
Comparison of NaI(Tl), CdTe, and HgI2 surgical probes: physical characterization
Abstract
The physical properties of three surgical probes containing different radiation detectors are compared: a NaI(Tl) scintillator with a flexible, fiber-optic light guide, and two semiconductor detectors that operate at room temperature, CdTe and HgI2. Also compared are spectra, energy resolutions, and counting efficiencies measured at a variety of gamma-ray energies between 30 and 1000 keV. The energy resolution of the NaI probe is substantially poorer than that of either semiconductor probe due in part to light losses in coupling the scintillator to the fiber optics. The semiconductor probes have complex spectral response due to charge-carrier trapping and K x-ray escape, and not all photoelectric interactions in these detectors contribute to the useful part of the photopeak. Above 120 keV the counting efficiency for the NaI probe is an order of magnitude higher than for the CdTe and HgI2 probes. Both energy resolution and counting efficiency are slightly better for the HgI2 probe than for the CdTe probe.
Similar articles
-
Comparison of NaI(T1), CdTe, and HgI2 surgical probes: effect of scatter compensation on probe performance.Med Phys. 1991 May-Jun;18(3):382-9. doi: 10.1118/1.596683. Med Phys. 1991. PMID: 1870479
-
Solid state detectors in nuclear medicine.Q J Nucl Med. 2002 Mar;46(1):3-7. Q J Nucl Med. 2002. PMID: 12072840 Review.
-
An evaluation of HgI2 detectors for x-ray computed tomography.Med Phys. 1980 Jul-Aug;7(4):370-3. doi: 10.1118/1.594694. Med Phys. 1980. PMID: 7393164
-
NaI gamma camera performance for high energies: Effects of crystal thickness, photomultiplier tube geometry and light guide thickness.Med Phys. 2024 Jul;51(7):4696-4708. doi: 10.1002/mp.17043. Epub 2024 Apr 3. Med Phys. 2024. PMID: 38569052
-
The intraoperative gamma probe: basic principles and choices available.Semin Nucl Med. 2000 Jan;30(1):33-48. doi: 10.1016/s0001-2998(00)80060-4. Semin Nucl Med. 2000. PMID: 10656242 Review.
Cited by
-
Gamma probes and their use in tumor detection in colorectal cancer.Int Semin Surg Oncol. 2008 Nov 19;5:25. doi: 10.1186/1477-7800-5-25. Int Semin Surg Oncol. 2008. PMID: 19019238 Free PMC article.
-
Plasma exudation in the skin measured by external detection of conversion electrons.Eur J Nucl Med. 1996 Mar;23(3):290-4. doi: 10.1007/BF00837627. Eur J Nucl Med. 1996. PMID: 8599960
-
Feasibility evaluation of radioimmunoguided surgery of breast cancer.Int J Mol Imaging. 2012;2012:545034. doi: 10.1155/2012/545034. Epub 2012 Feb 20. Int J Mol Imaging. 2012. PMID: 22518303 Free PMC article.
-
A comprehensive overview of radioguided surgery using gamma detection probe technology.World J Surg Oncol. 2009 Jan 27;7:11. doi: 10.1186/1477-7819-7-11. World J Surg Oncol. 2009. PMID: 19173715 Free PMC article. Review.
Publication types
MeSH terms
Substances
Grants and funding
LinkOut - more resources
Full Text Sources
Other Literature Sources