Ionization chamber volume determination and quality assurance using micro-CT imaging
- PMID: 18723930
- DOI: 10.1088/0031-9155/53/18/012
Ionization chamber volume determination and quality assurance using micro-CT imaging
Abstract
Ion chamber dosimetry requires a high degree of precision, at all steps within the dosimetric process, in order to ensure accurate dose measurements. This work presents a novel technique for ion chamber volume determination and quality assurance, using micro-computed tomography (micro-CT). Four nominally identical Exradin A1SL chambers (0.056 cm(3)) (Standard Imaging, WI, USA) were imaged using a micro-CT system (GE Locus, GE Healthcare, London, Ontario) and irradiated in a 6 MV x-ray reference field. Air volumes were calculated from the CT datasets using 3D analysis software (Microview 2.1.1, General Electric Healthcare, London, Ontario). Differences in the volumes of each chamber determined using micro-CT images agreed with differences in the ionization response within 1% for each chamber. Calibration coefficients were then compared through cross-calibration with a calibrated ion chamber and from the CT-measured volumes. The average ratio of these values was found to be 0.958 +/- 0.009 indicating good correlation. The results demonstrate the promise of using micro-CT imaging for the absolute volumetric characterization of ion chambers. The images have the potential to be an important clinical tool for quality assurance of ion chamber construction and integrity after routine clinical usage.
Comment in
-
Comments on 'Ionization chamber volume determination and quality assurance using micro-CT imaging'.Phys Med Biol. 2009 Mar 21;54(6):L23-7; author reply L29-30. doi: 10.1088/0031-9155/54/6/L01. Epub 2009 Feb 19. Phys Med Biol. 2009. PMID: 19229094
Similar articles
-
Characterization and use of a 2D-array of ion chambers for brachytherapy dosimetric quality assurance.Med Dosim. 2012 Autumn;37(3):250-6. doi: 10.1016/j.meddos.2011.09.004. Epub 2011 Dec 19. Med Dosim. 2012. PMID: 22189031
-
A patient-specific quality assurance study on absolute dose verification using ionization chambers of different volumes in RapidArc treatments.Med Dosim. 2012 Winter;37(4):436-41. doi: 10.1016/j.meddos.2012.04.001. Epub 2012 May 23. Med Dosim. 2012. PMID: 22626968
-
Comments on 'Ionization chamber volume determination and quality assurance using micro-CT imaging'.Phys Med Biol. 2009 Mar 21;54(6):L23-7; author reply L29-30. doi: 10.1088/0031-9155/54/6/L01. Epub 2009 Feb 19. Phys Med Biol. 2009. PMID: 19229094
-
CT radiation dose and image quality.Radiol Clin North Am. 2005 Nov;43(6):953-62, vii. doi: 10.1016/j.rcl.2005.07.002. Radiol Clin North Am. 2005. PMID: 16253656 Review.
-
Metrology for ultrasonic applications.Prog Biophys Mol Biol. 2007 Jan-Apr;93(1-3):138-52. doi: 10.1016/j.pbiomolbio.2006.07.023. Epub 2006 Aug 10. Prog Biophys Mol Biol. 2007. PMID: 17081597 Review.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical