Ionization chamber volume determination and quality assurance using micro-CT imaging
- PMID: 18723930
- DOI: 10.1088/0031-9155/53/18/012
Ionization chamber volume determination and quality assurance using micro-CT imaging
Abstract
Ion chamber dosimetry requires a high degree of precision, at all steps within the dosimetric process, in order to ensure accurate dose measurements. This work presents a novel technique for ion chamber volume determination and quality assurance, using micro-computed tomography (micro-CT). Four nominally identical Exradin A1SL chambers (0.056 cm(3)) (Standard Imaging, WI, USA) were imaged using a micro-CT system (GE Locus, GE Healthcare, London, Ontario) and irradiated in a 6 MV x-ray reference field. Air volumes were calculated from the CT datasets using 3D analysis software (Microview 2.1.1, General Electric Healthcare, London, Ontario). Differences in the volumes of each chamber determined using micro-CT images agreed with differences in the ionization response within 1% for each chamber. Calibration coefficients were then compared through cross-calibration with a calibrated ion chamber and from the CT-measured volumes. The average ratio of these values was found to be 0.958 +/- 0.009 indicating good correlation. The results demonstrate the promise of using micro-CT imaging for the absolute volumetric characterization of ion chambers. The images have the potential to be an important clinical tool for quality assurance of ion chamber construction and integrity after routine clinical usage.
Comment in
-
Comments on 'Ionization chamber volume determination and quality assurance using micro-CT imaging'.Phys Med Biol. 2009 Mar 21;54(6):L23-7; author reply L29-30. doi: 10.1088/0031-9155/54/6/L01. Epub 2009 Feb 19. Phys Med Biol. 2009. PMID: 19229094
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical