Early follow-up for a hybrid total hip arthroplasty using a metal-backed acetabular component designed to reduce "backside" polyethylene wear
- PMID: 18751806
- PMCID: PMC2504131
- DOI: 10.1007/s11420-005-0102-6
Early follow-up for a hybrid total hip arthroplasty using a metal-backed acetabular component designed to reduce "backside" polyethylene wear
Abstract
Modular acetabular systems have several advantages; however, increased polyethylene wear has been recognized as a disadvantage. In our study, an acetabular shell design with a highly polished inner surface, a locking mechanism that minimizes micromotion, and a high polyethylene liner to shell conformity was evaluated. A prospective cohort of 50 consecutive hips was followed up for a mean of 3.6 years (range, 2-7 years). One patient required an early revision for recurrent instability. Mean linear head penetration rates were 0.242 mm/year (range, 0.048-0.655 mm/year). The true linear wear rate was calculated after subtracting the linear wear in the first follow-up year from the total wear at the end of the follow-up. The true linear wear rate was 0.173 mm/year (range, 0.03-0.423 mm/year). A positive correlation was found between wear rate and male gender. The low head penetration rates suggest that a polished liner with an improved locking mechanism and increased liner conformity can reduce polyethylene wear.
Similar articles
-
Effect of acetabular modularity on polyethylene wear and osteolysis in total hip arthroplasty.J Bone Joint Surg Am. 2002 Jan;84(1):58-63. doi: 10.2106/00004623-200201000-00009. J Bone Joint Surg Am. 2002. PMID: 11792780
-
Analysis of temporal wear patterns of porous-coated acetabular components: distinguishing between true wear and so-called bedding-in.J Bone Joint Surg Am. 1999 Jun;81(6):821-30. doi: 10.2106/00004623-199906000-00009. J Bone Joint Surg Am. 1999. PMID: 10391547
-
Current Total Knee Designs: Does Baseplate Roughness or Locking Mechanism Design Affect Polyethylene Backside Wear?Clin Orthop Relat Res. 2017 Dec;475(12):2970-2980. doi: 10.1007/s11999-017-5494-3. Epub 2017 Sep 13. Clin Orthop Relat Res. 2017. PMID: 28905208 Free PMC article.
-
[Research progress of backside wear in acetabular liners].Zhongguo Xiu Fu Chong Jian Wai Ke Za Zhi. 2013 Dec;27(12):1453-6. Zhongguo Xiu Fu Chong Jian Wai Ke Za Zhi. 2013. PMID: 24640364 Review. Chinese.
-
Cementing constrained acetabular liners in revision hip replacement: clinical and laboratory observations.Instr Course Lect. 2004;53:131-40. Instr Course Lect. 2004. PMID: 15116607 Review.
Cited by
-
Early polyethylene wear and excessive acetabular granuloma in an uncemented HA-coated total hip arthroplasty--midterm results of a prospective study.HSS J. 2006 Sep;2(2):114-20. doi: 10.1007/s11420-006-9001-8. HSS J. 2006. PMID: 18751822 Free PMC article.
References
-
- Brooker AF, Bowerman JW, Robinson RA, Riley LH. Ectopic ossification following total hip replacement. J Bone Joint Surg Am. 1973;55:1629–1632. - PubMed
-
- Bueche MJ, Herzenberg JE, Stubbs BT. Dissociation of a metal-backed polyethylene acetabular component. A case report. J Arthroplasty. 1989;4:39–41. - PubMed
-
- Callaghan JJ, Pedersen DR, Olejniczak JP, Goetz DD, Johnston RC. Radiographic measurement of wear in 5 cohorts of patients observed for 5 to 22 years. Clin Orthop. 1995;317:14–18. - PubMed
-
- Charnley J, Halley DK (1975) Rate of wear in total hip replacement. Clin Orthop 112:170–179 - PubMed
LinkOut - more resources
Full Text Sources