Pain and complications of directional vacuum-assisted stereotactic biopsy: comparison of the Mammotome and Vacora techniques
- PMID: 18755562
- DOI: 10.1016/j.ejrad.2008.07.015
Pain and complications of directional vacuum-assisted stereotactic biopsy: comparison of the Mammotome and Vacora techniques
Abstract
Objective: The aim of our study is to compare the Mammotome and Vacora methods of stereotactic directional vacuum-assisted biopsy in terms of pain and complications.
Materials and methods: From June 2001 to May 2005, 1114 consecutive patients underwent directional stereotactic vacuum-assisted breast biopsy (DVAB) for nonpalpable mammographically detected breast lesions (BI-RADS 3, 4 or 5). Respectively 967 and 147 patients underwent the Mammotome and Vacora procedures. Pain was evaluated with a visual analog scale. Immediate and late complications were recorded.
Results: The mean+/-S.D. (range) pain scores in the Mammotome and Vacora groups were 1.7+/-1.8 (0-9) and 2.9+/-2.3 (0-10), respectively (p<0.001). Patient age and operator experience were the main determinants of pain. Immediate complications were significantly more frequent in the Mammotome group (p=0.003), and so were late hematomas (p=0.04). Moderate and severe complications occurred exclusively in the Mammotome group.
Conclusion: The Mammotome technique is associated with a higher risk of immediate and late complications, while the Vacora technique is associated with more frequent severe pain. Patient age was the major factor influencing pain. Further prospective studies are needed to clarify factors incriminated in pain or complications after DVAB procedures.
Similar articles
-
Percutaneous excisional biopsy of clinically benign breast lesions with vacuum-assisted system: comparison of three devices.Eur J Radiol. 2012 Apr;81(4):725-30. doi: 10.1016/j.ejrad.2011.01.059. Epub 2011 Feb 5. Eur J Radiol. 2012. PMID: 21300503
-
Interventional bleeding, hematoma and scar-formation after vacuum-biopsy under stereotactic guidance: Mammotome(®)-system 11 g/8 g vs. ATEC(®)-system 12 g/9 g.Eur J Radiol. 2012 May;81(5):e739-45. doi: 10.1016/j.ejrad.2012.01.033. Epub 2012 Feb 29. Eur J Radiol. 2012. PMID: 22381441
-
Vacuum assisted stereotactic guided mammotome biopsies in the management of screen detected microcalcifications: experience of a large breast screening centre.J Clin Pathol. 2008 Jun;61(6):766-9. doi: 10.1136/jcp.2007.054130. Epub 2008 Mar 6. J Clin Pathol. 2008. PMID: 18326021
-
The use of a vacuum-assisted biopsy device (Mammotome) in the early detection of breast cancer in the United Arab Emirates.Ann N Y Acad Sci. 2008 Sep;1138:108-13. doi: 10.1196/annals.1414.016. Ann N Y Acad Sci. 2008. PMID: 18837890
-
[The epideiology of postoperative pain].Ann Fr Anesth Reanim. 1998;17(6):534-9. doi: 10.1016/s0750-7658(98)80038-5. Ann Fr Anesth Reanim. 1998. PMID: 9750792 Review. French.
Cited by
-
Feasibility and safety of image-guided vacuum-assisted breast biopsy: A PRISMA-compliant systematic review and meta-analysis of 20 000 population from 36 longitudinal studies.Int Wound J. 2019 Dec;16(6):1506-1512. doi: 10.1111/iwj.13224. Epub 2019 Sep 18. Int Wound J. 2019. PMID: 31531950 Free PMC article.
-
Adjunctive imprint cytology of core needle biopsy specimens improved diagnostic accuracy for breast cancer.Springerplus. 2013 Aug 6;2:372. doi: 10.1186/2193-1801-2-372. eCollection 2013. Springerplus. 2013. PMID: 24010031 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical