. 2009 Jan;38(1):51-9.
doi: 10.1007/s00256-008-0577-6. Epub 2008 Sep 5.

Diffusion tensor imaging and fiber tractography of the median nerve at 1.5T: optimization of b value

Affiliations

Diffusion tensor imaging and fiber tractography of the median nerve at 1.5T: optimization of b value

Gustav Andreisek et al. Skeletal Radiol. 2009 Jan.

Abstract

Objective: The objective of this study was to systematically assess the optimal b value for diffusion tensor imaging and fiber tractography of the median nerve at 1.5 T.

Materials and methods: This is a prospective study which was carried out with institutional review board approval and written informed consent from the study subjects. Fifteen healthy volunteers (seven men, eight women; mean age, 31.2 years) underwent diffusion tensor imaging of the wrist. A single-shot spin-echo-based echo-planar imaging sequence (TR/TE, 7000/103 ms) was performed in each subject at eight different b values ranging from 325 to 1,550 s/mm(2). Number and length of reconstructed fiber tracts, fiber density index (FDi), fractional anisotropy (FA), and apparent diffusion coefficient (ADC) were calculated for the median nerve. Signal-to-noise ratio (SNR) was also calculated for each acquisition. The overall image quality was assessed by two readers in consensus by ranking representative fiber tract images for each subject using a scale range from 1 to 8 (1 = best to 8 = worst image quality).

Results: Longest fibers were observed for b values between 675 and 1,025 s/mm(2). Maximum FDi was found at b values of 1,025 s/mm(2). FA was between 0.5 and 0.6 for all b values. ADC gradually decreased from 1.44 x 10(-3) to 0.92 x 10(-3) mm(2)/s with increasing b values. Maximum SNR +/- standard deviation (175.4 +/- 72.6) was observed at the lowest b value and decreased with increasing b values. SNR at b values of 1,025 s/mm(2) was 48.5% of the maximum SNR. Optimal fiber tract image quality was found for b values of 1,025 s/mm(2).

Conclusions: The optimal b value for diffusion tensor imaging and fiber tractography of the median nerve at 1.5 T was 1,025 s/mm(2).

PubMed Disclaimer

Similar articles

Cited by

References

    1. Anat Rec. 1999 Jun 15;257(3):102-9 - PubMed
    1. Magn Reson Imaging. 2007 Apr;25(3):406-11 - PubMed
    1. Magn Reson Med. 2001 Mar;45(3):448-53 - PubMed
    1. Radiology. 2003 Sep;228(3):603-8 - PubMed
    1. Magn Reson Med. 2004 Feb;51(2):230-6 - PubMed

Publication types

LinkOut - more resources