Evaluation of a continuous glucose monitoring system for home-use conditions
- PMID: 18777789
Evaluation of a continuous glucose monitoring system for home-use conditions
Abstract
Objective: To evaluate the safety and effectiveness of the FreeStyle Navigator Continuous Glucose Monitoring System when used by adult patients with type 1 or 2 diabetes requiring insulin in the home environment.
Research design and methods: Multicenter, prospective study enrolling 137 subjects. This was a two-phase study consisting of 20 days of use ("masked period") without access to continuous glucose readings followed by 21 days of use ("unmasked period") with access to continuous glucose readings and glucose alarms. At the conclusion of the study, all subjects were asked to complete a User's Questionnaire.
Results: A total of 11,487 paired continuous glucose sensor and blood glucose reference results from 961 sensors were evaluated; 77.2 percent were in Clarke error grid zone A and 19.6 percent were in zone B. Only 13.1 percent of the hypoglycemic and 0.5 percent of the hyperglycemic threshold alarms were false alerts. Subjects with type 1 diabetes demonstrated a 55 percent reduction in time spent with significant hypoglycemia (below 55 mg/dL) (P < .0001) from masked to unmasked periods. Their average number of hypoglycemic episodes (below 70 mg/dL) per day fell from 1.1 to 0.8 (P < .0001). Results from patient questionnaires demonstrated high levels of subject satisfaction and the ability to use and understand the system.
Conclusions: FreeStyle Navigator Continuous Glucose Monitoring System is safe and effective, and results in a high level of subject satisfaction while used in the home setting. Utilization of continuous glucose monitors with alerts may result in cost offsets by reducing the number and severity of clinically significant events and assisting in the maintenance of optimal glycemic control.
Similar articles
-
Accuracy of the 5-day FreeStyle Navigator Continuous Glucose Monitoring System: comparison with frequent laboratory reference measurements.Diabetes Care. 2007 May;30(5):1125-30. doi: 10.2337/dc06-1602. Epub 2007 Mar 2. Diabetes Care. 2007. PMID: 17337488
-
Changes in the glycemic profiles of women with type 1 and type 2 diabetes during pregnancy.Diabetes Care. 2007 Nov;30(11):2785-91. doi: 10.2337/dc07-0500. Epub 2007 Jul 31. Diabetes Care. 2007. PMID: 17666464 Clinical Trial.
-
Pilot evaluation of continuous subcutaneous glucose monitoring in children with multiple organ dysfunction syndrome.Pediatr Crit Care Med. 2010 May;11(3):415-9. doi: 10.1097/PCC.0b013e3181c59144. Pediatr Crit Care Med. 2010. PMID: 19924024
-
Sensor-augmented pump therapy in type 1 diabetes.Curr Opin Endocrinol Diabetes Obes. 2008 Apr;15(2):118-22. doi: 10.1097/MED.0b013e3282f7960b. Curr Opin Endocrinol Diabetes Obes. 2008. PMID: 18316945 Review.
-
[Biosensors for continuous glucose and lactate monitoring].Ned Tijdschr Geneeskd. 2003 Jun 21;147(25):1204-8. Ned Tijdschr Geneeskd. 2003. PMID: 12848054 Review. Dutch.
Cited by
-
Reducing glycaemic variability in type 1 diabetes self-management with a continuous glucose monitoring system based on wired enzyme technology.Diabetologia. 2009 Aug;52(8):1496-503. doi: 10.1007/s00125-009-1408-6. Epub 2009 Jun 13. Diabetologia. 2009. PMID: 19526212
-
Clinical implications and economic impact of accuracy differences among commercially available blood glucose monitoring systems.J Diabetes Sci Technol. 2013 Mar 1;7(2):365-80. doi: 10.1177/193229681300700213. J Diabetes Sci Technol. 2013. PMID: 23566995 Free PMC article.
-
Stillbirth: prevention and supportive bereavement care.BMJ Med. 2023 Jun 27;2(1):e000262. doi: 10.1136/bmjmed-2022-000262. eCollection 2023. BMJ Med. 2023. PMID: 37564829 Free PMC article. Review.
-
Clinical Performance Evaluation of Continuous Glucose Monitoring Systems: A Scoping Review and Recommendations for Reporting.J Diabetes Sci Technol. 2023 Nov;17(6):1506-1526. doi: 10.1177/19322968231190941. Epub 2023 Aug 20. J Diabetes Sci Technol. 2023. PMID: 37599389 Free PMC article.
-
A Critical Discussion of Alert Evaluations in the Context of Continuous Glucose Monitoring System Performance.J Diabetes Sci Technol. 2024 Jul;18(4):847-856. doi: 10.1177/19322968241236504. Epub 2024 Mar 13. J Diabetes Sci Technol. 2024. PMID: 38477308 Free PMC article. Review.
Publication types
MeSH terms
Substances
LinkOut - more resources
Medical