The relationship between programmed pacemaker pulse amplitude and the surface electrocardiogram recorded amplitude: application of a new high-bandwidth electrocardiogram system
- PMID: 18804786
- DOI: 10.1016/j.jelectrocard.2008.06.023
The relationship between programmed pacemaker pulse amplitude and the surface electrocardiogram recorded amplitude: application of a new high-bandwidth electrocardiogram system
Abstract
Background: Recording and displaying outputs from electronic pacemakers with electrocardiogram (ECG) recorders typically used in clinical practice have presented a number of technical limitations. We have recently reported on a new high-bandwidth ECG system and have shown that it is capable of reproducing accurate pulse amplitudes and durations from the body surface. In the present work, we have used our data to calculate a transform function between the programmed pacemaker output voltage and the amplitude on the body surface.
Methods: We recorded 3 high-bandwidth (75,000 samples per second) ECGs from each of 100 pacemaker patients at 3 different programmed outputs. Each pacemaker pulse was isolated using the criterion standard annotations, and the pulses were transformed from the 8 independent leads to an XYZ vector using the Dower transform. The magnitude of the vector was calculated. Linear regression techniques were used to learn a transfer function over the records of the first 50 patients. These results were tested against the second 50 patients.
Results: The measured pacemaker pulse vector magnitude has a linear relationship to the programmed pacemaker amplitude on a per-patient basis for most of the patients in the training database. The linear transform models were tested against the testing set with an R(2) metric of 0.38 for the atrial pulses and 0.54 for the right ventricular pulses.
Conclusion: Understanding the relationship between the generated pacemaker pulses and the measurements at the body surface will help drive specifications for pacemaker pulse detection among the various device manufactures.
Similar articles
-
Implanted cardiac pacemaker pulses as recorded from the body surface.J Electrocardiol. 2012 Nov-Dec;45(6):663-9. doi: 10.1016/j.jelectrocard.2012.08.008. Epub 2012 Oct 1. J Electrocardiol. 2012. PMID: 23034404
-
Observations of pacemaker pulses in high-bandwidth electrocardiograms and Dower-estimated vectorcardiograms.J Electrocardiol. 2011 Mar-Apr;44(2):275-81. doi: 10.1016/j.jelectrocard.2010.09.014. Epub 2010 Dec 3. J Electrocardiol. 2011. PMID: 21130466
-
Manifestation of left atrial events and interatrial frequency gradients in the surface electrocardiogram during atrial fibrillation: contributions from posterior leads.J Cardiovasc Electrophysiol. 2009 Nov;20(11):1231-6. doi: 10.1111/j.1540-8167.2009.01523.x. Epub 2009 Jun 26. J Cardiovasc Electrophysiol. 2009. PMID: 19563359
-
Diagnostic value of the 12-lead electrocardiogram during conventional and biventricular pacing for cardiac resynchronization.Cardiol Clin. 2006 Aug;24(3):471-90, x. doi: 10.1016/j.ccl.2006.05.001. Cardiol Clin. 2006. PMID: 16939837 Review.
-
Electronic pacemakers.Emerg Med Clin North Am. 2006 Feb;24(1):179-94, vii. doi: 10.1016/j.emc.2005.08.011. Emerg Med Clin North Am. 2006. PMID: 16308119 Review.
Cited by
-
A Novel High-Resolution Surface Electrocardiographic Method to Identify and Characterize Myocardial Scar: A Proof-of-Concept Study.CJC Open. 2021 May 27;3(10):1207-1213. doi: 10.1016/j.cjco.2021.05.010. eCollection 2021 Oct. CJC Open. 2021. PMID: 34888504 Free PMC article.
-
A Contactless Sensor for Pacemaker Pulse Detection: Design Hints and Performance Assessment.Sensors (Basel). 2018 Aug 18;18(8):2715. doi: 10.3390/s18082715. Sensors (Basel). 2018. PMID: 30126178 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Medical