Radiation protection to the eye and thyroid during diagnostic cerebral angiography: a phantom study

Affiliations

Radiation protection to the eye and thyroid during diagnostic cerebral angiography: a phantom study

C P Shortt et al. J Med Imaging Radiat Oncol. 2008 Aug.

Abstract

We measured radiation doses to the eye and thyroid during diagnostic cerebral angiography to assess the effectiveness of bismuth and lead shields at dose reduction. Phantom head angiographic studies were performed with bismuth (study 1) and lead shields (study 2). In study 1 (12 phantoms), thermoluminescent dosimeters (TLD) were placed over the eyes and thyroid in three groups: (i) no shields (four phantoms); (ii) anterior bismuth shields (four phantoms) and (iii) anterior and posterior bismuth shields (four phantoms). In a second study (eight phantoms), lead shields were placed over the thyroid only and TLD dose measurements obtained in two groups: (i) no shielding (four phantoms) and (ii) thyroid lead shielding (four phantoms). A standard 4-vessel cerebral angiogram was performed on each phantom. Study 1 (bismuth shields) showed higher doses to the eyes compared with thyroid (mean 13.03 vs 5.98 mSv, P < 0.001) and a higher eye dose on the X-ray tube side. Overall, the use of bismuth shielding did not significantly reduce dose to either eyes or thyroid in the measured TLD positions. In study 2, a significant thyroid dose reduction was found with the use of lead shields (47%, mean 2.46 vs 4.62 mSv, P < 0.001). Considerable doses to the eyes and thyroid highlight the need for increased awareness of patient protection. Eye shielding is impractical and interferes with diagnostic capability. Thyroid lead shielding yields significant protection to the thyroid, is not in the field of view and should be used routinely.