Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2009 Mar;19(3):591-8.
doi: 10.1007/s00330-008-1181-9. Epub 2008 Sep 26.

ACCURATUM: improved calcium volume scoring using a mesh-based algorithm--a phantom study

Affiliations

ACCURATUM: improved calcium volume scoring using a mesh-based algorithm--a phantom study

Stefan C Saur et al. Eur Radiol. 2009 Mar.

Abstract

To overcome the limitations of the classical volume scoring method for quantifying coronary calcifications, including accuracy, variability between examinations, and dependency on plaque density and acquisition parameters, a mesh-based volume measurement method has been developed. It was evaluated and compared with the classical volume scoring method for accuracy, i.e., the normalized volume (measured volume/ground-truthed volume), and for variability between examinations (standard deviation of accuracy). A cardiac computed-tomography (CT) phantom containing various cylindrical calcifications was scanned using different tube voltages and reconstruction kernels, at various positions and orientations on the CT table and using different slice thicknesses. Mean accuracy for all plaques was significantly higher (p < 0.0001) for the proposed method (1.220 +/- 0.507) than for the classical volume score (1.896 +/- 1.095). In contrast to the classical volume score, plaque density (p = 0.84), reconstruction kernel (p = 0.19), and tube voltage (p = 0.27) had no impact on the accuracy of the developed method. In conclusion, the method presented herein is more accurate than classical calcium scoring and is less dependent on tube voltage, reconstruction kernel, and plaque density.

PubMed Disclaimer

Similar articles

Cited by

References

    1. J Am Coll Cardiol. 1990 Mar 15;15(4):827-32 - PubMed
    1. Radiology. 2003 Jul;228(1):101-6 - PubMed
    1. IEEE Trans Med Imaging. 2007 Mar;26(3):273-82 - PubMed
    1. Radiology. 2002 Oct;225(1):113-9 - PubMed
    1. AJR Am J Roentgenol. 2000 Mar;174(3):803-9 - PubMed

Publication types