Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Comparative Study
. 2008 Dec;17(12):1757-65.
doi: 10.1007/s00586-008-0797-4. Epub 2008 Oct 8.

Biomechanical comparison of two different concepts for stand alone anterior lumbar interbody fusion

Affiliations
Comparative Study

Biomechanical comparison of two different concepts for stand alone anterior lumbar interbody fusion

Philipp Schleicher et al. Eur Spine J. 2008 Dec.

Abstract

Segmental instability in degenerative disc disease is often treated with anterior lumbar interbody fusion (ALIF). Current techniques require an additional posterior approach to achieve sufficient stability. The test device is an implant which consists of a PEEK-body and an integrated anterior titanium plate hosting four diverging locking screws. The test device avoids posterior fixation by enhancing stability via the locking screws. The test device was compared to an already established stand alone interbody implant in a human cadaveric three-dimensional stiffness test. In the biomechanical test, the L4/5 motion segment of 16 human cadaveric lumbar spines were isolated and divided into two test groups. Tests were performed in flexion, extension, right and left lateral bending, right and left axial rotation. Each specimen was tested in native state first, then a discectomy was performed and either of the test implants was applied. Finite element analysis (FE) was also performed to investigate load and stress distribution within the implant in several loading conditions. The FE models simulated two load cases. These were flexion and extension with a moment of 5 Nm. The biomechanical testing revealed a greater stiffness in lateral bending for the SynFix-LR compared to the established implant. Both implants showed a significantly higher stiffness in all loading directions compared to the native segment. In flexion loading, the PEEK component takes on most of the load, whereas the majority of the extension load is put on the screws and the screw-plate junction. Clinical investigation of the test device seems reasonable based on the good results reported here.

PubMed Disclaimer

Figures

Fig. 1
Fig. 1
SynFix; anterocaudal view with locking screws inserted
Fig. 2
Fig. 2
STALIF; anterior view with two of four possible screws inserted
Fig. 3
Fig. 3
The meshed model of vertebra for analysis of pure axial compression and tension contains 224,183 elements
Fig. 4
Fig. 4
The meshed model of cage, plate and screw for analysis of pure axial compression and tension contains 321,586 elements
Fig. 5
Fig. 5
Relative results of range of motion (ROM) measurements. Native segment is set as 1 (red line). The black bar indicates the median value, the edges of the box indicate first and third percentile, respectively. Error bars indicate maximum/minimum values. Outliers are defined as a data point extending greater than 1.5 box-lengths from the edge of a boxplot, extremes are defined as data points which extend greater than three box lengths from the edge of a boxplot. Significant differences between groups are indicated
Fig. 6
Fig. 6
Relative results of neutral zone (NZ) measurements. Native segment is set as 1
Fig. 7
Fig. 7
Oblique reconstruction showing (a) maximal tensile stress and (b) von-Mises comparison stress distribution of the cage, stabilization plate and fixation screw when tested with a flexion load of 5 Nm
Fig. 8
Fig. 8
Oblique reconstruction showing (a) maximal tensile stress and (b) von-Mises comparison stress distribution of the cage, stabilization plate and fixation screw when tested with an extension load of 5 Nm

References

    1. Arand M, Wilke HJ, Schultheiss M, Hartwig E, Kinzl L, Claes L. Comparative stability of the “Internal Fixator” and the “Universal Spine System” and the effect of crosslinking transfixating systems. A biomechanical in vitro study. Biomed Tech (Berl) 2000;45:311–316. doi: 10.1515/bmte.2000.45.11.311. - DOI - PubMed
    1. Bozkus H, Chamberlain RH, Perez Garza LE, Crawford NR, Dickman CA. Biomechanical comparison of anterolateral plate, lateral plate, and pedicle screws-rods for enhancing anterolateral lumbar interbody cage stabilization. Spine. 2004;29:635–641. doi: 10.1097/01.BRS.0000115126.13081.7D. - DOI - PubMed
    1. Brox JI, Sorensen R, Friis A, Nygaard O, Indahl A, Keller A, et al. Randomized clinical trial of lumbar instrumented fusion and cognitive intervention and exercises in patients with chronic low back pain and disc degeneration. Spine. 2003;28:1913–1921. doi: 10.1097/01.BRS.0000083234.62751.7A. - DOI - PubMed
    1. Cain CM, Schleicher P, Gerlach R, Pflugmacher R, Scholz M, Kandziora F. A new stand-alone anterior lumbar interbody fusion device: biomechanical comparison with established fixation techniques. Spine. 2005;30:2631–2636. doi: 10.1097/01.brs.0000187897.25889.54. - DOI - PubMed
    1. Charriere E, Sirey F, Zysset PK. A finite element model of the L5–S1 functional spinal unit: development and comparison with biomechanical tests in vitro. Comput Methods Biomech Biomed Engin. 2003;6:249–261. doi: 10.1080/10255840310001606099. - DOI - PubMed

Publication types

MeSH terms

LinkOut - more resources