Rates and severity of perforation from implantable cardioverter-defibrillator leads: a 4-year study

Affiliations
Controlled Clinical Trial

Rates and severity of perforation from implantable cardioverter-defibrillator leads: a 4-year study

Mintu Turakhia et al. J Interv Card Electrophysiol. 2009 Jan.

Abstract

Introduction: Although recent case reports and one small single-year observational study report a substantially increased rate of perforation with the St. Jude Riata series defibrillator lead, these results have not been externally validated.

Methods and results: From 2004 to 2007, 593 implantable cardioverter-defibrillator (ICD) implants were performed by six faculty and 13 fellows at four University of California, San Francisco, CA, USA and affiliate hospitals. An electronic medical records system was systematically searched to identify clinically significant cases of ICD lead dislodgment or perforation. Of 307 (56%) St. Jude leads (all Riata series 6.3- and 7.3-French leads), 188 (29%) Medtronic leads (including 99 Sprint Fidelis 6.6 French leads), and 98 (15%) Guidant/Boston Scientific leads, there were three perforations in 593 cases (0.51%). One perforation occurred with a Medtronic Sprint Fidelis 6949 lead (0.53%), and two with a St. Jude Riata 1581 lead (0.65%). There were no statistically significant differences in perforation or dislodgement rates between manufacturers or lead models (p = NS for all). In both cases of perforation with the St. Jude Riata leads, the lead tip perforated through the pericardium into the pleural space.

Conclusions: In our 4-year series of ICD implants, perforation and dislodgement rates were low, similar across all lead makes and models, and well below published and accepted complication rates. Our findings contradict previously-reported higher rates of perforation with the Riata lead. Registry and product performance reports should also classify complications by severity and outcome to provide a more complete assessment of product safety.

PubMed Disclaimer

References

    1. Pacing Clin Electrophysiol. 2008 Jan;31(1):1-2 - PubMed
    1. Heart Rhythm. 2007 Apr;4(4):439-42 - PubMed
    1. Heart Rhythm. 2007 Sep;4(9):1248; author reply 1248 - PubMed
    1. Pacing Clin Electrophysiol. 2008 Jan;31(1):13-5 - PubMed
    1. Heart Rhythm. 2007 Sep;4(9):1248-9; author reply 1249 - PubMed

Publication types

LinkOut - more resources