Steam pops during irrigated radiofrequency ablation: feasibility of impedance monitoring for prevention

Affiliations

Steam pops during irrigated radiofrequency ablation: feasibility of impedance monitoring for prevention

Jens Seiler et al. Heart Rhythm. 2008 Oct.

Abstract

Background: Steam pops are a risk of irrigated radiofrequency catheter ablation (RFA) and may cause cardiac perforation. Data to guide radiofrequency (RF) energy titration to avoid steam pops are limited.

Objective: This study sought to assess the frequency and consequence of audible pops and to determine the feasibility of using the magnitude of impedance change to predict pops.

Methods: We reviewed consecutive endocardial open-irrigated RFA for ventricular tachycardia (VT) with continuously recorded ablation data in 142 patients with structural heart disease. Steam pops were defined as an audible pop associated with a sudden spike in impedance. Ablation lesions before or after pops served as controls.

Results: From a total of 4,107 ablation lesions, 62 (1.5%) steam pops occurred in 42 procedures in 38 patients. Perforation with tamponade occurred with 1 of 62 (2%) pops. Applications with pops had a greater impedance decrease (22 +/- 7 Omega vs. 18 +/- 8 Omega, P = .001) and a higher maximum power (45 +/- 5 W vs. 43 +/- 6 W, P = .011), but did not differ in maximum catheter tip temperature (40 degrees C +/- 4 degrees C vs. 40 degrees C +/- 4 degrees C, P = .180) from applications without pops. Eighty percent of pops occurred after impedance decreased by at least 18 Omega.

Conclusion: During VT ablation with open irrigation, audible pops are infrequent and do not usually cause perforation. Limiting RF power to achieve an impedance decrease of <18 Omega is a feasible method of reducing the likelihood of a pop when perforation risk is of concern.