A conductance sensor for dissolved sulphur dioxide using a series piezoelectric crystal device
Abstract
A new piezoelectric crystal impedance sensor for the determination of sulphur dioxide in aqueous solution is presented. It is realized using a series piezoelectric crystal device which is constructed by connecting an AT-cut piezoelectric crystal to a probe in series. The probe is filled with an internal electrolyte solution that is separated from sample solutions by a gas-permeable membrane. The present sensor exhibits a favourable frequency response to 1 x 10(-7)-1 x 10(-3) M sulphur dioxide. The detection limit is 1 x 10(-8) M. The effects of the sensor preparation are considered. Dynamic range, reproducibility, response time and selectivity of the sensor are also discussed. The proposed sensor has been used successfully for lamp sulphur determinations in petroleum samples.
LinkOut - more resources
Full Text Sources