Characterization and use of a Raman liquid-core waveguide sensor using preconcentration principles
- PMID: 18968968
- DOI: 10.1016/S0039-9140(02)00623-9
Characterization and use of a Raman liquid-core waveguide sensor using preconcentration principles
Abstract
A novel Raman sensor using a liquid-core optical waveguide is reported, implementing a Teflon-AF 2400 tube filled with water. An aqueous analyte mixture of benzene, toluene and p-xylene was introduced using a 1000 microl sample loop to the liquid-core waveguide (LCW) sensor and the analytes were preconcentrated on the inside surface of the waveguide tubing. The analytes were then eluted from the waveguide using an acetonitrile-water solvent mixture injected via a 30 microl eluting solvent loop. The preconcentration factor was experimentally determined to be 14-fold, in reasonable agreement with the theoretical preconcentration factor of 33 based upon the sample volume to elution volume ratio. Raman spectra of benzene, toluene and p-xylene were obtained during elution. It was found that analytically useful Raman signals for benzene, toluene and p-xylene were obtained at 992, 1004 and 1206 cm(-1), respectively. The relative standard deviation of the method was 3% for three replicate measurements. The limit of detection (LOD) was determined to be 730 ppb (parts per billion by volume) for benzene, exceptional for a system that does not resort to surface enhancement or resonance Raman approaches. The Raman spectra of these test analytes were evaluated for qualitative and quantitative analysis utility.
Similar articles
-
Quantitative analysis of Raman signal enhancement from aqueous samples in liquid core optical fibers.Appl Spectrosc. 2004 Oct;58(10):1165-71. doi: 10.1366/0003702042336109. Appl Spectrosc. 2004. PMID: 15527516
-
Raman sensitivity enhancement for aqueous absorbing sample using Teflon-AF 2400 liquid core optical fibre cell.Anal Chim Acta. 2007 Jan 2;581(1):154-8. doi: 10.1016/j.aca.2006.07.082. Epub 2006 Aug 7. Anal Chim Acta. 2007. PMID: 17386439
-
Absorbance characteristics of a liquid-phase gas sensor based on gas-permeable liquid core waveguides.Anal Chim Acta. 2013 Dec 4;804:207-14. doi: 10.1016/j.aca.2013.10.020. Epub 2013 Oct 18. Anal Chim Acta. 2013. PMID: 24267083
-
Simultaneous quantitative determination of benzene, toluene, and xylenes in water using mid-infrared evanescent field spectroscopy.Anal Chem. 2004 May 1;76(9):2643-8. doi: 10.1021/ac0347009. Anal Chem. 2004. PMID: 15117210
-
On-line sequential injection dispersive liquid-liquid microextraction system for flame atomic absorption spectrometric determination of copper and lead in water samples.Talanta. 2009 Jun 30;79(1):86-91. doi: 10.1016/j.talanta.2009.03.005. Epub 2009 Mar 14. Talanta. 2009. PMID: 19376348
Cited by
-
Characterization of a novel ultra low refractive index material for biosensor application.Sens Actuators B Chem. 2009 Aug 18;141(1):227-232. doi: 10.1016/j.snb.2009.04.034. Sens Actuators B Chem. 2009. PMID: 20161155 Free PMC article.
LinkOut - more resources
Full Text Sources