Abstract
Objectives: To identify an effective bioassay for human hepatitis B virus (HBV) using voltammetric sensor.
Study design: Electrode was prepared with bovine IgG immobilized onto a DNA-linked carbon nanotube electrode (BIDCE). Analytical BIDCE conditions were optimized through square wave (SW) stripping and cyclic voltammetry (CV).
Results: Optimum parameters of 75 Hz SW frequency, 0.1 V SW amplitude, -1.3 V SW accumulation potential, 100 s accumulation time, and 3.0 mV SW increment potential were obtained. Within working ranges of 0.035-0.242 mg/mL anti-bovine IgG, the relative standard deviation of 0.2 mL HBV was observed to be 0.04 (n=4).
Conclusions: Diagnostic application was performed through direct assay of HBV in non-treated human blood.
Similar articles
-
Diagnosis of copper ions in vascular tracts using a fluorine-doped carbon nanotube sensor.Talanta. 2008 Feb 15;74(5):1635-41. doi: 10.1016/j.talanta.2007.10.017. Epub 2007 Oct 16. Talanta. 2008. PMID: 18371829
-
Detection of dopamine in the pharmacy with a carbon nanotube paste electrode using voltammetry.Bioelectrochemistry. 2006 May;68(2):227-31. doi: 10.1016/j.bioelechem.2005.09.002. Epub 2005 Nov 23. Bioelectrochemistry. 2006. PMID: 16309972
-
Diagnosis of Helicobacter pylori bacterial infections using a voltammetric biosensor.J Microbiol Methods. 2011 Oct;87(1):44-8. doi: 10.1016/j.mimet.2011.07.002. Epub 2011 Jul 23. J Microbiol Methods. 2011. PMID: 21801757
-
Biosensors for hepatitis B virus detection.World J Gastroenterol. 2014 Sep 21;20(35):12485-92. doi: 10.3748/wjg.v20.i35.12485. World J Gastroenterol. 2014. PMID: 25253948 Free PMC article. Review.
-
Diagnosis of hepatitis B virus infection through serological and virological markers.Expert Rev Gastroenterol Hepatol. 2008 Aug;2(4):553-62. doi: 10.1586/17474124.2.4.553. Expert Rev Gastroenterol Hepatol. 2008. PMID: 19072403 Review.
Cited by
-
Ultrasensitive Label-Free Sensing of IL-6 Based on PASE Functionalized Carbon Nanotube Micro-Arrays with RNA-Aptamers as Molecular Recognition Elements.Biosensors (Basel). 2017 Apr 17;7(2):17. doi: 10.3390/bios7020017. Biosensors (Basel). 2017. PMID: 28420169 Free PMC article.
-
Carbon Nanotube Field-Effect Transistor-Based Chemical and Biological Sensors.Sensors (Basel). 2021 Feb 2;21(3):995. doi: 10.3390/s21030995. Sensors (Basel). 2021. PMID: 33540641 Free PMC article. Review.
-
Biosensors for Detection of Hepatitis B Virus: Review.Curr Med Chem. 2025;32(23):4725-4741. doi: 10.2174/0109298673295474240531100312. Curr Med Chem. 2025. PMID: 38860910 Review.
-
Development of a Novel, Ultra-rapid Biosensor for the Qualitative Detection of Hepatitis B Virus-associated Antigens and Anti-HBV, Based on "Membrane-engineered" Fibroblast Cells with Virus-Specific Antibodies and Antigens.Sensors (Basel). 2009;9(3):2176-86. doi: 10.3390/s90302176. Epub 2009 Mar 25. Sensors (Basel). 2009. PMID: 22574007 Free PMC article.
-
Decadal Journey of CNT-Based Analytical Biosensing Platforms in the Detection of Human Viruses.Nanomaterials (Basel). 2022 Nov 23;12(23):4132. doi: 10.3390/nano12234132. Nanomaterials (Basel). 2022. PMID: 36500755 Free PMC article. Review.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical