Effects of flow amplitudes on intraprong pressures during bubble versus ventilator-generated nasal continuous positive airway pressure in premature infants
- PMID: 18977980
- DOI: 10.1542/peds.2007-3416
Effects of flow amplitudes on intraprong pressures during bubble versus ventilator-generated nasal continuous positive airway pressure in premature infants
Abstract
Objectives: The goal were to characterize the flow dependence of bubble nasal continuous positive airway pressure delivery in a cohort of preterm infants and to compare the actual (delivered) intraprong continuous positive airway pressure with the intended (set) nasal continuous positive airway pressure for both ventilator-generated nasal continuous positive airway pressure and bubble nasal continuous positive airway pressure delivery. A range of set values and constant flow rates were studied in the same preterm infants.
Methods: For 12 premature infants of <1500 g (birth weight: 1140 +/- 267 g; gestational age: 28.5 +/- 1.9 weeks; study age: 12.9 +/- 8 days; all mean +/- SD), intraprong pressures were measured at 3 increasing flow settings, repeated for set nasal continuous positive airway pressures (or desired immersion depths) of 4 and 6 cm H(2)O. Next, intraprong pressures were measured at bubble nasal continuous positive airway pressure expiratory tubing submersion depths and ventilator-generated nasal continuous positive airway pressure set expiratory pressures of 2, 3, 4, 5, and 7 cm H(2)O while the flow rate was held constant.
Results: Actual (delivered) intraprong pressure during bubble nasal continuous positive airway pressure delivery was highly flow dependent and increased as the flow rate increased. During ventilator-generated nasal continuous positive airway pressure delivery, actual pressure at the nasal prongs closely approximated the pressure set at the ventilator. During bubble nasal continuous positive airway pressure delivery at constant flow rate, the average delivered prong pressure was 1.3 cm H(2)O (range: 0.5-2.2 cm H(2)O) higher than that set through submersion of the expiratory tubing, and the relative difference between the set and actual pressures increased at lesser immersion depths.
Conclusions: Prong pressure during bubble nasal continuous positive airway pressure delivery is highly variable and depends on the interaction of submersion depth and flow amplitudes.
Similar articles
-
A randomized controlled trial of post-extubation bubble continuous positive airway pressure versus Infant Flow Driver continuous positive airway pressure in preterm infants with respiratory distress syndrome.J Pediatr. 2009 May;154(5):645-50. doi: 10.1016/j.jpeds.2008.12.034. Epub 2009 Feb 23. J Pediatr. 2009. PMID: 19230906 Clinical Trial.
-
Pressure variation during ventilator generated nasal intermittent positive pressure ventilation in preterm infants.Arch Dis Child Fetal Neonatal Ed. 2010 Sep;95(5):F359-64. doi: 10.1136/adc.2009.172957. Epub 2010 May 20. Arch Dis Child Fetal Neonatal Ed. 2010. PMID: 20488862
-
Bubble and ventilator-derived nasal continuous positive airway pressure in premature infants: work of breathing and gas exchange.J Perinatol. 2011 Jan;31(1):44-50. doi: 10.1038/jp.2010.55. Epub 2010 Apr 15. J Perinatol. 2011. PMID: 20393478 Clinical Trial.
-
Continuous positive airway pressure: scientific and clinical rationale.Curr Opin Pediatr. 2008 Apr;20(2):119-24. doi: 10.1097/MOP.0b013e3282f63953. Curr Opin Pediatr. 2008. PMID: 18332704 Review.
-
The nursing care of the infant receiving bubble CPAP therapy.Adv Neonatal Care. 2008 Apr;8(2):78-95; quiz 96-7. doi: 10.1097/01.ANC.0000317256.76201.72. Adv Neonatal Care. 2008. PMID: 18418205 Review.
Cited by
-
A Fixed Flow is More Effective than Titrated Flow during Bubble Nasal CPAP for Respiratory Distress in Preterm Neonates.Front Pediatr. 2015 Oct 12;3:81. doi: 10.3389/fped.2015.00081. eCollection 2015. Front Pediatr. 2015. PMID: 26528456 Free PMC article.
-
Mechanisms of nasal high flow therapy in newborns.J Appl Physiol (1985). 2020 Apr 1;128(4):822-829. doi: 10.1152/japplphysiol.00871.2019. Epub 2020 Feb 20. J Appl Physiol (1985). 2020. PMID: 32078463 Free PMC article.
-
The Effect of Inhaled Budesonide on the Prevention of Chronic Lung Disease in Premature Neonates with Respiratory Distress Syndrome.Int J Prev Med. 2018 Feb 8;9:15. doi: 10.4103/ijpvm.IJPVM_336_16. eCollection 2018. Int J Prev Med. 2018. PMID: 29541430 Free PMC article.
-
Safety and effectiveness of bubble continuous positive airway pressure in preterm neonates with respiratory distress.Med J Armed Forces India. 2014 Oct;70(4):327-31. doi: 10.1016/j.mjafi.2013.08.003. Epub 2014 Sep 26. Med J Armed Forces India. 2014. PMID: 25382905 Free PMC article.
-
Bubble vs conventional continuous positive airway pressure for prevention of extubation failure in preterm very low birth weight infants: a pilot study.Indian J Pediatr. 2012 Sep;79(9):1163-8. doi: 10.1007/s12098-011-0651-2. Epub 2011 Dec 23. Indian J Pediatr. 2012. PMID: 22189698 Clinical Trial.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical