Objective: We report the first case of clip obliteration of a cerebral aneurysm containing a displaced microstent.
Clinical presentation: A 63-year-old woman presented with a 6-month history of headaches. She had no other neurological symptoms, and her examination was normal. A computed tomographic scan showed no evidence of hemorrhage. Angiography confirmed the presence of a large left paraclinoid aneurysm.
Intervention: The patient initially underwent attempted endovascular repair of the aneurysm. A Neuroform microstent (Boston Scientific, Natick, MA), which was placed across the aneurysm neck, migrated into the aneurysm upon passing through it with the microcatheter to be used to place coils. This left the proximal half of the stent in the aneurysm and the distal half in the internal carotid artery. Attempts to remove the stent and to place a second stent across the neck of the aneurysm failed, and any plans to place coils were aborted. Three months after the failed endovascular procedure, the patient consented to a craniotomy. The aneurysm was successfully obliterated by placing several clips directly across the neck of the aneurysm and leaving the stent in place.
Conclusion: The patient experienced an excellent clinical and angiographic outcome. Although not the ideal treatment strategy, this report illustrates that clip obliteration of aneurysms containing displaced microstents can be performed successfully for this complication of endovascular treatment.
Similar articles
-
Thrombosis of a fusiform intracranial aneurysm induced by overlapping neuroform stents: case report.Neurosurgery. 2007 May;60(5):E950-1; discussion E950-1. doi: 10.1227/01.NEU.0000255427.08926.DC. Neurosurgery. 2007. PMID: 17460508
-
Stent remodeling technique for coiling of ruptured wide-neck cerebral aneurysms: case report.Neurosurgery. 2009 Nov;65(5):E1007-8; discussion E1008. doi: 10.1227/01.NEU.0000350984.86323.61. Neurosurgery. 2009. PMID: 19834390
-
Novel use of an endovascular embolectomy device for retrograde suction decompression-assisted clip ligation of a large paraclinoid aneurysm: technical case report.Neurosurgery. 2008 May;62(5 Suppl 2):ONSE412-3; discussion ONSE413-4. doi: 10.1227/01.neu.0000326027.39467.88. Neurosurgery. 2008. PMID: 18596535
-
Endovascular management of a ruptured cavernous carotid artery aneurysm associated with a carotid cavernous fistula with an intracranial self-expanding microstent and hydrogel-coated coil embolization: case report and review of the literature.Surg Neurol. 2007 Nov;68(5):562-7; discussion 567. doi: 10.1016/j.surneu.2006.10.074. Surg Neurol. 2007. PMID: 17961748 Review.
-
Novel surgical treatment of a transverse-sigmoid sinus aneurysm presenting as pulsatile tinnitus: technical case report.Neurosurgery. 2009 Feb;64(2):E393-4; discussion E394. doi: 10.1227/01.NEU.0000338255.93266.00. Neurosurgery. 2009. PMID: 19190447 Review.
Cited by
-
Neuroform-EZ Stent Modification for Facilitated Delivery during Intracranial Aneurysm Embosurgery.Neurointervention. 2013 Sep;8(2):101-4. doi: 10.5469/neuroint.2013.8.2.101. Epub 2013 Aug 29. Neurointervention. 2013. PMID: 24024074 Free PMC article.
-
Endovascular repositioning of a pipeline embolization device dislocated from the vertebral into the basilar artery using a stent-in-stent technique. Practical and technical considerations.Clin Neuroradiol. 2012 Mar;22(1):47-54. doi: 10.1007/s00062-011-0128-8. Clin Neuroradiol. 2012. PMID: 22286147
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical