Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2009 Jan;73(1):43-55.
doi: 10.1016/j.ijporl.2008.09.023. Epub 2008 Nov 4.

Air-conduction estimated from tympanometry (ACET): 2. The use of hearing level-ACET discrepancy (HAD) to determine appropriate use of bone-conduction tests in identifying permanent and mixed impairments

Collaborators

Air-conduction estimated from tympanometry (ACET): 2. The use of hearing level-ACET discrepancy (HAD) to determine appropriate use of bone-conduction tests in identifying permanent and mixed impairments

Mark Haggard et al. Int J Pediatr Otorhinolaryngol. 2009 Jan.

Abstract

Objective: The caseload at secondary care in paediatric otology is largely otitis media with effusion (OME) and highly recurrent acute otitis media (RAOM). Few of these cases merit suspicion for hearing loss beyond the middle ear. The companion paper showed that the air conduction estimated from tympanometry (ACET) formula, derived on a very large clinical sample referred for ear or hearing problems and pre-assessed for a clinical trial, gives usable although only approximate estimates for hearing level (HL) on such a caseload. Tympanometry corresponds to a conductive loss (i.e. air-bone gap) so the HL-ACET discrepancy (HAD) should approximate the bone-conduction (BC) threshold. Clinical criteria might enable HAD to substitute for BC tests where those are infeasible, or to identify those most needing BC tests.

Method: ACET had been derived for the 4-frequency binaural average on 3085 cases with tympanometry and air-conduction HL available. On the 2701 of those with BC data at 1kHz, we re-calculated ACET for 1kHz only, and then explored the sensitivity/specificity trade of the discrepancy (HAD) in detecting clinically significant BC levels and the correlation between these measures. We further illustrated the generalization of the formula and cut-off on a small separate retrospective clinical sample.

Results: Correlations were moderate in the clinically relevant region. There were five cases of BC> or =30dB in the database. At a HAD cut-off of +5dB, the sift would identify all (nominal 100% sensitivity). For marginal cases, two definitions were adopted (BC> or =25dB and > or =20dB; 9 and 23 cases, respectively). Sift sensitivity remained high (89% and 83%, respectively), and specificity was acceptable (75% for both definitions).

Conclusions: Given tympanometry and air-conduction HL, comparison of HAD with a recommended cut-off gives acceptable sensitivity and specificity for non-OME hearing problems. BC testing can be reserved for probable positive cases, provisionally only 25% of caseload. HAD could temporarily substitute for BC measurement in children too young to accept bone-conduction transducers in awake testing. Where a high proportion of the caseload is expected to have middle ear fluid, ACET and HAD together offer efficient possibilities for best use of available information.

PubMed Disclaimer

LinkOut - more resources