Dual-source CT coronary angiography in patients with atrial fibrillation: comparison with single-source CT
- PMID: 18993003
- DOI: 10.1016/j.ejrad.2008.09.011
Dual-source CT coronary angiography in patients with atrial fibrillation: comparison with single-source CT
Abstract
Objective: To evaluate the performance of dual-source computed tomography (DSCT) for the visualization of the coronary arteries in a population with atrial fibrillation (AF) compared to single-source CT (SSCT) and to explore the impact of patients' heart rate (HR) on image quality (IQ) and reconstruction timing.
Methods: Thirty consecutive patients (11 male, 19 female; 69.0+/-9.2 years old) with suspected coronary artery disease and permanent AF were examined on a DSCT scanner (120 kV, 400 mAs/rot, 0.33 s rotation time, 64 x 0.6 mm collimation, pitch 0.20-0.28, Siemens Somatom Definition). Patients were divided into two groups: low and medium HR group (HR<or=80 bpm, n=14) and high HR group (HR>80 bpm, n=16). Five of the patients also underwent conventional coronary angiography (CAG). The raw data from both tube detector arrays were reconstructed as DSCT data using a routine algorithm (temporal resolution of 83 ms). The raw data from one tube detector array was reconstructed separately on the same system using a routine single source algorithm (temporal resolution of 83-165 ms) and defined as virtual SSCT data. Image quality was assessed using a four-point grading scale from excellent (1) to non-assessable (4).
Results: IQ of the DSCT data was significantly better than that of the virtual SSCT data (mean score 1.33+/-0.61 vs. 1.80+/-1.02; Z=-8.755, P=0.000). 98.6% of the segments shown in DSCT were diagnostic, compared with 89.9% of the segments in virtual SSCT, chi(2)=32.595, P=0.000. In DSCT group, IQ of low HR group was also better than that of high HR group, although the difference was not as big (mean score 1.25+/-0.52 vs. 1.38+/-0.66; Z=-2.227, P=0.026). The mean HR of low HR group and high HR group were 67.4+/-8.5 beats per minute (bpm) and 94.2+/-8.8 bpm (t=-8.499, P=0.000). The range of the variation of HR was higher in high HR group than in low HR group (mean difference between maximum and minimum HR 79.5+/-21.0 vs. 49.9+/-21.1 bpm; t=-3.845, P=0.001). In 23 (77%) patients optimal IQ was achieved within one phase for all three main arteries. In low HR group, the optimal phase was distributed evenly between diastole and systole; but in high HR group, the optimal phase shifted to systole in most cases (92%). In five cases these results were compared to CAG results to look for the ability to identify stenosis with a diameter reduction of more than 50% of the lumen. With DSCT it was possible to diagnose 20 of 21 stenosis and 48 of 49 non-stenosed vessel segments correctly. With SSCT 19 of 21 stenosis and 45 of 49 non-stenosed vessel segments were diagnosed correctly according to CAG.
Conclusion: Due to its high temporal resolution, DSCT provides images of full diagnostic image quality in patients with AF, which otherwise would be excluded from CT examinations. The number of assessable segments for DSCT is still somewhat less than in non-AF patients, but in opposition to SSCT still valid for routine diagnostic imaging. Patients' HR had impact on IQ and reconstruction timing.
Similar articles
-
Noninvasive coronary angiography using dual-source computed tomography in patients with atrial fibrillation.Invest Radiol. 2009 Mar;44(3):159-67. doi: 10.1097/RLI.0b013e3181948b05. Invest Radiol. 2009. PMID: 19151607
-
Prospective-triggered sequential dual-source end-systolic coronary CT angiography for patients with atrial fibrillation: a feasibility study.J Cardiovasc Comput Tomogr. 2013 Mar-Apr;7(2):102-9. doi: 10.1016/j.jcct.2013.02.002. Epub 2013 Feb 27. J Cardiovasc Comput Tomogr. 2013. PMID: 23545461
-
Dose performance and image quality: dual source CT versus single source CT in cardiac CT angiography.Eur J Radiol. 2009 Dec;72(3):396-400. doi: 10.1016/j.ejrad.2008.08.010. Epub 2008 Oct 2. Eur J Radiol. 2009. PMID: 18835119
-
Current and Novel Imaging Techniques in Coronary CT.Radiographics. 2015 Jul-Aug;35(4):991-1010. doi: 10.1148/rg.2015140181. Epub 2015 Jun 5. Radiographics. 2015. PMID: 26046942 Review.
-
[The clinical value of computer tomography (CT) of diagnostics of acute thorax pain--a literature review].Przegl Lek. 2013;70(3):123-7. Przegl Lek. 2013. PMID: 24003665 Review. Polish.
Cited by
-
What are the basic concepts of temporal, contrast, and spatial resolution in cardiac CT?J Cardiovasc Comput Tomogr. 2009 Nov-Dec;3(6):403-8. doi: 10.1016/j.jcct.2009.07.003. Epub 2009 Jul 30. J Cardiovasc Comput Tomogr. 2009. PMID: 19717355 Free PMC article. Review.
-
Prospectively ECG-Triggered Sequential Dual-Source Coronary CT Angiography in Patients with Atrial Fibrillation: Influence of Heart Rate on Image Quality and Evaluation of Diagnostic Accuracy.PLoS One. 2015 Jul 29;10(7):e0134194. doi: 10.1371/journal.pone.0134194. eCollection 2015. PLoS One. 2015. PMID: 26221952 Free PMC article.
-
Controlling rate and rhythm increases feasibility of CT angiography in atrial fibrillation.Clin Med Res. 2010 Jul;8(2):96-8. doi: 10.3121/cmr.2010.874. Clin Med Res. 2010. PMID: 20660933 Free PMC article.
-
Assessment of Cardiac Lead Perforation: Comparison Among Chest Radiography, Transthoracic Echocardiography and Electrocardiography-gated Contrast-enhanced Cardiac CT.Eur Radiol. 2019 Feb;29(2):963-974. doi: 10.1007/s00330-018-5633-6. Epub 2018 Jul 17. Eur Radiol. 2019. PMID: 30019144
-