Effects of a 2-, 3- and 4-electrode stimulator design on current dispersion on the surface and into the limb during electrical stimulation in controls and patients with wounds
- PMID: 19005963
- DOI: 10.1080/03091900701574407
Effects of a 2-, 3- and 4-electrode stimulator design on current dispersion on the surface and into the limb during electrical stimulation in controls and patients with wounds
Abstract
Electrical stimulation is a widely used modality in the field of physical therapy and exercise physiology. The most common method for the application of electrical stimulation is a two-electrode system where one electrode is the source and the other is a reference. However, recent studies report that a more effective delivery system can be achieved if more than two electrodes are used. In the present investigation, the circuitry to deliver electrical stimulation through a 2-, 3- or 4-electrode delivery system was designed. The system was evaluated by its ability to deliver current on the surface of the skin as well as deep into the quadriceps muscle in six control subjects and in and around wounds in six other subjects. The results of the experiments showed that much better depth of penetration was achieved in a 4-electrode system (one electrode was on the opposite side of the limb and three electrodes were on top of the limb) than in either a 2- or a 3-electrode delivery system. In non-wounded skin, given the same current from the stimulator, the current in the quadriceps muscle was found to be double with a 4-electrode versus a 2-electrode system. In wounds, this same finding was seen. Here, blood flow, an indicator of the effectiveness of electrical stimulation in wounds, was three times higher if a multi-channel stimulator was used versus a 2-channel stimulator. Thus a multi-channel electrical stimulation system is more effective than a 2-electrode system.
Similar articles
-
A multi-channel stimulator and electrode array providing a rotating current whirlpool for electrical stimulation of wounds.J Med Eng Technol. 2008 Sep-Oct;32(5):371-84. doi: 10.1080/03091900601116994. J Med Eng Technol. 2008. PMID: 18821415
-
Current distribution under electrodes in relation to stimulation current and skin blood flow: are modern electrodes really providing the current distribution during stimulation we believe they are?J Med Eng Technol. 2006 Nov-Dec;30(6):368-81. doi: 10.1080/03091900500183855. J Med Eng Technol. 2006. PMID: 17060165
-
A re-evaluation of modelling of the current flow between electrodes: consideration of blood flow and wounds.J Med Eng Technol. 2007 Jan-Feb;31(1):62-74. doi: 10.1080/03091900600687698. J Med Eng Technol. 2007. PMID: 17365428 Clinical Trial.
-
Using the nerve stimulator for peripheral or plexus nerve blocks.Minerva Anestesiol. 2006 Jun;72(6):467-71. Minerva Anestesiol. 2006. PMID: 16682917 Review.
-
Electrical stimulation to restore respiration.J Rehabil Res Dev. 1996 Apr;33(2):123-32. J Rehabil Res Dev. 1996. PMID: 8724168 Review.
Cited by
-
Electrical Stimulation Therapy and Wound Healing: Where Are We Now?Adv Wound Care (New Rochelle). 2012 Dec;1(6):238-243. doi: 10.1089/wound.2011.0351. Adv Wound Care (New Rochelle). 2012. PMID: 24527312 Free PMC article. Review.
-
Estimation of the distribution of intramuscular current during electrical stimulation of the quadriceps muscle.Eur J Appl Physiol. 2008 Jun;103(3):265-73. doi: 10.1007/s00421-008-0700-3. Epub 2008 Feb 23. Eur J Appl Physiol. 2008. PMID: 18297301 Clinical Trial.
-
Angiogenesis is induced and wound size is reduced by electrical stimulation in an acute wound healing model in human skin.PLoS One. 2015 Apr 30;10(4):e0124502. doi: 10.1371/journal.pone.0124502. eCollection 2015. PLoS One. 2015. PMID: 25928356 Free PMC article.
-
Electrical Stimulation and Cutaneous Wound Healing: A Review of Clinical Evidence.Healthcare (Basel). 2014 Oct 27;2(4):445-67. doi: 10.3390/healthcare2040445. Healthcare (Basel). 2014. PMID: 27429287 Free PMC article. Review.
MeSH terms
LinkOut - more resources
Full Text Sources