Comparative rotational stability of single-piece open-loop acrylic and plate-haptic silicone toric intraocular lenses
- PMID: 19006728
- DOI: 10.1016/j.jcrs.2008.07.012
Comparative rotational stability of single-piece open-loop acrylic and plate-haptic silicone toric intraocular lenses
Abstract
Purpose: To prospectively compare the early rotational stability of AcrySof SN60T toric intraocular lenses (IOLs) with that in a retrospective series of AA4203 toric IOLs.
Setting: Private practice, Los Altos, California, USA.
Methods: One hundred consecutive eyes with an AcrySof SN60T(3, 4, or 5) toric IOL were compared with a consecutive series of 90 AA4203 (TL or TF) toric IOLs. The same surgeon performed all IOL implantations using an identical surgical technique. In addition to deviation from the desired axis, the change in refractive cylinder was measured 1 month postoperatively.
Results: Although surgery was performed during different time periods, the 2 populations had a similar distribution of patient age, axial lengths, and spherical IOL powers. In the AcrySof SN60T group, 90%, 99%, and 100% of the IOLs were aligned at or within 5, 10, and 15 degrees, respectively, of the desired axis and in the AA4203 group, 70%, 90%, and 97%, respectively. The mean IOL rotation was 5.56 degrees +/- 8.49 (SD) in the AA4203 group and 3.35 +/- 3.41 degrees in the AcrySof SN60T group (P = .0232). One AcrySof SN60T IOL (1%) and 8 AA4203 IOLs (8.9%) were 15 degrees or more off axis (P = .01). No AcrySof SN60T IOL and 3.3% of AA4203 IOLs required surgical repositioning.
Conclusions: Both toric IOLs had good rotational stability and were effective in reducing preexisting corneal astigmatism. Based on the mean axis deviation and the number of IOLs rotating 15 degrees or more, the AcrySof SN60T toric IOL showed statistically better rotational stability.
Similar articles
-
Comparison of the visual outcomes and OPD-scan results of AMO Tecnis toric and Alcon Acrysof IQ toric intraocular lenses.J Refract Surg. 2012 Aug;28(8):551-5. doi: 10.3928/1081597X-20120703-03. Epub 2012 Jul 13. J Refract Surg. 2012. PMID: 22785061 Clinical Trial.
-
Foldable toric intraocular lens for astigmatism correction in cataract patients.J Cataract Refract Surg. 2008 Apr;34(4):601-7. doi: 10.1016/j.jcrs.2007.11.033. J Cataract Refract Surg. 2008. PMID: 18361982
-
Evaluation of the long-term rotational stability of single-piece, acrylic intraocular lenses.Br J Ophthalmol. 2010 Aug;94(8):1003-6. doi: 10.1136/bjo.2009.163485. Epub 2010 May 14. Br J Ophthalmol. 2010. PMID: 20472745 Clinical Trial.
-
Status of toric intraocular lenses.Curr Opin Ophthalmol. 2007 Feb;18(1):58-61. doi: 10.1097/ICU.0b013e328011f9bf. Curr Opin Ophthalmol. 2007. PMID: 17159449 Review.
-
A new slant on toric intraocular lens power calculation.J Refract Surg. 2013 May;29(5):348-54. doi: 10.3928/1081597X-20130415-06. J Refract Surg. 2013. PMID: 23659233 Review.
Cited by
-
Toric intraocular lens implantation in cataract patients with corneal opacity.BMC Ophthalmol. 2020 Mar 13;20(1):98. doi: 10.1186/s12886-020-01352-w. BMC Ophthalmol. 2020. PMID: 32169054 Free PMC article.
-
Factors Predictive of Refractive Error After Toric Lens Implantation.Clin Ophthalmol. 2023 Jun 26;17:1813-1821. doi: 10.2147/OPTH.S407414. eCollection 2023. Clin Ophthalmol. 2023. PMID: 37397954 Free PMC article.
-
Intraoperative Aberrometer Plus Image-guided System for Astigmatism Correction Compared to Standard Image-guided System for Significant Lens Opacity.In Vivo. 2025 Mar-Apr;39(2):1033-1041. doi: 10.21873/invivo.13908. In Vivo. 2025. PMID: 40010944 Free PMC article.
-
Rotational stability of monofocal and diffractive multifocal toric intraocular lens with identical design and material: a propensity score based prospective comparative study.BMC Ophthalmol. 2024 Feb 16;24(1):72. doi: 10.1186/s12886-024-03281-4. BMC Ophthalmol. 2024. PMID: 38365667 Free PMC article.
-
Cataract surgery with implantation of an artificial lens.Dtsch Arztebl Int. 2009 Oct;106(43):695-702. doi: 10.3238/arztebl.2009.0695. Epub 2009 Oct 23. Dtsch Arztebl Int. 2009. PMID: 19946433 Free PMC article. Review.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical
Miscellaneous