Central corneal thickness measured by the Orbscan II system, contact ultrasound pachymetry, and the Artemis 2 system
- PMID: 19006737
- PMCID: PMC2601704
- DOI: 10.1016/j.jcrs.2008.07.013
Central corneal thickness measured by the Orbscan II system, contact ultrasound pachymetry, and the Artemis 2 system
Abstract
Purpose: To compare central corneal thickness (CCT) measurements by the Orbscan II device, contact ultrasound (US) pachymetry, and the noncontact Artemis 2 scanning US system.
Setting: Department of Ophthalmology, Weill Cornell Medical College, New York, New York, USA.
Methods: The CCT in 40 eyes (20 normal subjects) was measured by the Orbscan II followed by contact US pachymetry and then the Artemis 2. Results were compared using analysis of variance (ANOVA), paired t tests, and Bland-Altman plots.
Results: There was a significant difference in CCT measurements between the 3 modes (F = 32.84, P = .0001, 1-way ANOVA). Artemis 2 and US pachymetry measurements were highly correlated (r2 = 0.963, P < .0001), although Artemis 2 values were a mean of 11.2 microm +/- 6.6 (SD) thinner than pachymetry values. Artemis 2 and Orbscan II measurements were less well correlated (r2 = 0.851, P < .001); Orbscan II values were a mean of 7.5 +/- 15.7 microm thinner than Artemis 2 values. Orbscan II values showed a trend toward increasing underestimation of CCT in thinner corneas.
Conclusions: Ultrasound pachymetry and Artemis 2 CCT measurements were highly correlated; the 11 microm mean difference in measurements may be attributed to decentration, oblique incidence of the probe to the cornea, or possibly the effect of topical anesthesia with contact pachymetry. Although the mean difference between Orbscan II and Artemis 2 values was 7.5 microm, Orbscan values were less correlated than Artemis 2 values with contact US pachymetry and were prone to underestimation of the CCT in thinner corneas.
Figures




Similar articles
-
Central corneal thickness measurements with partial coherence interferometry, ultrasound, and the Orbscan system.Ophthalmology. 2004 May;111(5):875-9. doi: 10.1016/j.ophtha.2003.09.027. Ophthalmology. 2004. PMID: 15121362
-
An assessment of regional differences in corneal thickness in normal human eyes, using the Orbscan II or ultrasound pachymetry.Optometry. 2007 Apr;78(4):181-90. doi: 10.1016/j.optm.2006.08.018. Optometry. 2007. PMID: 17400140
-
The reliability of central corneal thickness measurements by ultrasound and by Orbscan system in schoolchildren.Curr Eye Res. 2006 Jul-Aug;31(7-8):569-75. doi: 10.1080/02713680600801115. Curr Eye Res. 2006. PMID: 16877265
-
Comparison of central corneal thickness measurements by ultrasonic pachymetry, Orbscan II, and SP3000P in eyes with glaucoma or glaucoma suspect.Chang Gung Med J. 2012 May-Jun;35(3):255-62. doi: 10.4103/2319-4170.106146. Chang Gung Med J. 2012. PMID: 22735057
-
The orbscan acoustic (correction) factor for central corneal thickness measures of normal human corneas.Eye Contact Lens. 2010 Mar;36(2):106-15. doi: 10.1097/ICL.0b013e3181d0b604. Eye Contact Lens. 2010. PMID: 20107419 Review.
Cited by
-
Comparison of artemis 2 ultrasound and Visante optical coherence tomography corneal thickness profiles.J Refract Surg. 2013 Jan;29(1):36-41. doi: 10.3928/1081597X-20121126-01. Epub 2012 Nov 30. J Refract Surg. 2013. PMID: 23205905 Free PMC article.
-
Repeatability and interobserver reproducibility of Artemis-2 high-frequency ultrasound in determination of human corneal thickness.Clin Ophthalmol. 2012;6:761-9. doi: 10.2147/OPTH.S31690. Epub 2012 May 21. Clin Ophthalmol. 2012. PMID: 22693418 Free PMC article.
-
Repeatability of layered corneal pachymetry with the artemis very high-frequency digital ultrasound arc-scanner.J Refract Surg. 2010 Sep;26(9):646-59. doi: 10.3928/1081597X-20091105-01. Epub 2009 Nov 16. J Refract Surg. 2010. PMID: 19928698 Free PMC article.
-
Comparison of central corneal thickness measurements using ultrasound pachymetry, ultrasound biomicroscopy, and the Artemis-2 VHF scanner in normal eyes.Clin Ophthalmol. 2012;6:1037-43. doi: 10.2147/OPTH.S32955. Epub 2012 Jul 6. Clin Ophthalmol. 2012. PMID: 22848145 Free PMC article.
-
Central Corneal Thickness in Highly Myopic Eyes: Inter-device Agreement of Ultrasonic Pachymetry, Pentacam and Orbscan II Before and After Photorefractive Keratectomy.J Ophthalmic Vis Res. 2014 Jan;9(1):14-21. J Ophthalmic Vis Res. 2014. PMID: 24982727 Free PMC article.
References
-
- Bechman M, Thiel MJ, Neubauer AS, et al. Central corneal thickness measurements with a retinal optical coherence tomography device versus standard ultrasonic pachymetry. Cornea. 2001;20:50–54. - PubMed
-
- Bovelle R, Kaufman SC, Thompson HW, et al. Corneal thickness measurements with the Topcon SP-2000p specular microscope and an ultrasound pachymeter. Arch Ophthalmol. 1999;117:868–870. - PubMed
-
- Rainer G, Findl O, Petternel V, et al. Central corneal thickness measurements with partial coherence interferometry, ultrasound, and the Orbscan System. Ophthalmology. 2004;111:875–879. - PubMed
-
- Li HF, Petroll WM, Møller-Pederson T, et al. Epithelial and corneal thickness measurements by in vivo confocal microscopy through focusing (CMTF) Curr Eye Res. 1997;16:214–221. - PubMed
Publication types
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources
Medical