A randomized clinical trial to evaluate the safety and efficacy of a percutaneous left ventricular assist device versus intra-aortic balloon pumping for treatment of cardiogenic shock caused by myocardial infarction
- PMID: 19007597
- DOI: 10.1016/j.jacc.2008.05.065
A randomized clinical trial to evaluate the safety and efficacy of a percutaneous left ventricular assist device versus intra-aortic balloon pumping for treatment of cardiogenic shock caused by myocardial infarction
Abstract
Objectives: The aim of this study was to test whether the left ventricular assist device (LVAD) Impella LP2.5 (Abiomed Europe GmbH, Aachen, Germany) provides superior hemodynamic support compared with the intra-aortic balloon pump (IABP).
Background: Cardiogenic shock caused by left ventricular failure is associated with high mortality in patients with acute myocardial infarction (AMI). An LVAD may help to bridge patients to recovery from left ventricular failure.
Methods: In a prospective, randomized study, 26 patients with cardiogenic shock were studied. The primary end point was the change of the cardiac index (CI) from baseline to 30 min after implantation. Secondary end points included lactic acidosis, hemolysis, and mortality after 30 days.
Results: In 25 patients the allocated device (n = 13 IABP, n = 12 Impella LP2.5) could be safely placed. One patient died before implantation. The CI after 30 min of support was significantly increased in patients with the Impella LP2.5 compared with patients with IABP (Impella: DeltaCI = 0.49 +/- 0.46 l/min/m(2); IABP: DeltaCI = 0.11 +/- 0.31 l/min/m(2); p = 0.02). Overall 30-day mortality was 46% in both groups.
Conclusions: In patients presenting with cardiogenic shock caused by AMI, the use of a percutaneously placed LVAD (Impella LP 2.5) is feasible and safe, and provides superior hemodynamic support compared with standard treatment using an intra-aortic balloon pump. (Efficacy Study of LV Assist Device to Treat Patients With Cardiogenic Shock [ISAR-SHOCK]; NCT00417378).
Similar articles
-
Mechanical circulatory support with the Impella® LP5.0 pump and an intra-aortic balloon pump for cardiogenic shock in acute myocardial infarction: The IMPELLA-STIC randomized study.Arch Cardiovasc Dis. 2020 Apr;113(4):237-243. doi: 10.1016/j.acvd.2019.10.005. Epub 2019 Nov 15. Arch Cardiovasc Dis. 2020. PMID: 31740272 Clinical Trial.
-
Percutaneous left ventricular assist devices vs. intra-aortic balloon pump counterpulsation for treatment of cardiogenic shock: a meta-analysis of controlled trials.Eur Heart J. 2009 Sep;30(17):2102-8. doi: 10.1093/eurheartj/ehp292. Epub 2009 Jul 18. Eur Heart J. 2009. PMID: 19617601 Review.
-
Intra-Aortic Balloon Counterpulsation in Patients With Chronic Heart Failure and Cardiogenic Shock: Clinical Response and Predictors of Stabilization.J Card Fail. 2015 Nov;21(11):868-76. doi: 10.1016/j.cardfail.2015.06.383. Epub 2015 Jul 9. J Card Fail. 2015. PMID: 26164215 Free PMC article.
-
Impella Support for Acute Myocardial Infarction Complicated by Cardiogenic Shock.Circulation. 2019 Mar 5;139(10):1249-1258. doi: 10.1161/CIRCULATIONAHA.118.036614. Circulation. 2019. PMID: 30586755
-
Mechanical circulatory support with Impella versus intra-aortic balloon pump or medical treatment in cardiogenic shock-a critical appraisal of current data.Clin Res Cardiol. 2019 Nov;108(11):1249-1257. doi: 10.1007/s00392-019-01458-2. Epub 2019 Mar 21. Clin Res Cardiol. 2019. PMID: 30900010
Cited by
-
Pre-PCI versus immediate post-PCI Impella initiation in acute myocardial infarction complicated by cardiogenic shock.PLoS One. 2020 Jul 20;15(7):e0235762. doi: 10.1371/journal.pone.0235762. eCollection 2020. PLoS One. 2020. PMID: 32687502 Free PMC article. Clinical Trial.
-
Position Statement on Vascular Access Safety for Percutaneous Devices in AMI Complicated by Cardiogenic Shock.JACC Cardiovasc Interv. 2022 Oct 24;15(20):2003-2019. doi: 10.1016/j.jcin.2022.08.041. JACC Cardiovasc Interv. 2022. PMID: 36265932 Free PMC article. Review.
-
Intra-Aortic Balloon Pump and Ischemic Cardiogenic Shock May Still Be a Valuable Association.J Clin Med. 2021 Feb 16;10(4):778. doi: 10.3390/jcm10040778. J Clin Med. 2021. PMID: 33669179 Free PMC article.
-
Mechanical Support in Early Cardiogenic Shock: What Is the Role of Intra-aortic Balloon Counterpulsation?Curr Heart Fail Rep. 2020 Oct;17(5):247-260. doi: 10.1007/s11897-020-00480-0. Curr Heart Fail Rep. 2020. PMID: 32870448 Free PMC article. Review.
-
Comparative safety of percutaneous ventricular assist device and intra-aortic balloon pump in acute myocardial infarction-induced cardiogenic shock.Open Heart. 2021 Jun;8(1):e001662. doi: 10.1136/openhrt-2021-001662. Open Heart. 2021. PMID: 34127531 Free PMC article. Review.
Publication types
MeSH terms
Associated data
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical