Design and optimization of a high-speed, high-sensitivity, spinning disk confocal microscopy system
- PMID: 19021437
- DOI: 10.1117/1.2992141
Design and optimization of a high-speed, high-sensitivity, spinning disk confocal microscopy system
Abstract
We describe the principles, design, and systems integration of a flexible, high-speed, high-sensitivity, high-resolution confocal spinning disk microscopy (SDCM) system. We present several artifacts unique to high-speed SDCM along with techniques to minimize them. We show example experimental results from a specific implementation capable of generating 3-D image stacks containing 30 2-D slices at 30 stacks per second. This implementation also includes optics for differential interference contrast (DIC), phase, and bright-field imaging, as well as an optical trap with sensitive force and position measurement.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources