Frequency of instrument breakage during orthopaedic procedures and its effects on patients
- PMID: 19047710
- DOI: 10.2106/JBJS.H.00163
Frequency of instrument breakage during orthopaedic procedures and its effects on patients
Abstract
Background: While breakage of an orthopaedic instrument is a relatively rare occurrence, orthopaedic surgeons need to be familiar with this complication and how to deal with it. Relatively little information about this subject has been published.
Methods: Every case of instrument breakage during orthopaedic procedures performed in two hospitals during a two-year period was documented prospectively. All patients were followed for a postoperative period ranging from twelve to thirty-six months, during which radiographs in two planes were made to assess changes in, or migration of, the broken object.
Results: During the observation period, 11,856 surgical procedures were performed in the two hospitals. The overall rate of instrument breakage was 0.35%. The broken piece was removed in five cases, and the broken instrument was left in situ in thirty-seven cases. During the follow-up period, none of the patients had any symptoms.
Conclusions: In most cases, breakage of an orthopaedic instrument is not a problem. Any instance of instrument breakage should be fully documented in the surgical report.
Similar articles
-
The rate of instrument breakage during orthopaedic procedures.Int Orthop. 2002;26(3):185-7. doi: 10.1007/s00264-002-0344-6. Epub 2002 Mar 26. Int Orthop. 2002. PMID: 12073114 Free PMC article.
-
Drill bit failure without implant involvement--an intraoperative complication in orthopaedic surgery.Injury. 1992;23 Suppl 2:S5-16. doi: 10.1016/s0020-1383(10)80002-6. Injury. 1992. PMID: 1446952
-
Foreign-body aspiration during microlaryngoscopy: an unusual case of instrument failure.J Pediatr Surg. 2007 Feb;42(2):e13-4. doi: 10.1016/j.jpedsurg.2006.10.023. J Pediatr Surg. 2007. PMID: 17270532
-
A multiinstitutional review of central venous line complications: retained intravascular fragments.J Pediatr Surg. 2009 May;44(5):972-6. doi: 10.1016/j.jpedsurg.2009.01.033. J Pediatr Surg. 2009. PMID: 19433181 Review.
-
Retained central venous lines (CVLs) after attempted removal: an 11-year series and literature review.J Pediatr Surg. 2013 Sep;48(9):1887-91. doi: 10.1016/j.jpedsurg.2013.01.050. J Pediatr Surg. 2013. PMID: 24074662 Review.
Cited by
-
Power-Tool Use in Orthopaedic Surgery: Iatrogenic Injury, Its Detection, and Technological Advances: A Systematic Review.JB JS Open Access. 2021 Nov 19;6(4):e21.00013. doi: 10.2106/JBJS.OA.21.00013. eCollection 2021 Oct-Dec. JB JS Open Access. 2021. PMID: 34841185 Free PMC article.
-
Occult risk of broken instruments for endoscopy-assisted surgery.World J Surg. 2014 Nov;38(11):3015-22. doi: 10.1007/s00268-014-2666-0. World J Surg. 2014. PMID: 24952076
-
[Reportable incidents with surgical instruments in orthopedic surgery].Orthopade. 2014 Jun;43(6):561-7. doi: 10.1007/s00132-014-2296-0. Orthopade. 2014. PMID: 24832376 Review. German.
-
Breakage and detachment of the rigid cystoscope's distal tip: an unusual case of urological instrument malfunction.Cent European J Urol. 2024;77(2):334-338. doi: 10.5173/ceju.2024.20. Epub 2024 Mar 31. Cent European J Urol. 2024. PMID: 39345320 Free PMC article.
-
Successful retrieval of a broken pituitary rongeur tip from the lumbar intervertebral disc space using a magnetic probe: illustrative case.J Neurosurg Case Lessons. 2024 Dec 2;8(23):CASE24377. doi: 10.3171/CASE24377. Print 2024 Dec 2. J Neurosurg Case Lessons. 2024. PMID: 39622041 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical