Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Clinical Trial
. 2009 Mar;10(2):227-33.
doi: 10.1097/PCC.0b013e3181937227.

Extravascular lung water measurement using transpulmonary thermodilution in children

Affiliations
Clinical Trial

Extravascular lung water measurement using transpulmonary thermodilution in children

Joris Lemson et al. Pediatr Crit Care Med. 2009 Mar.

Abstract

Objective: Measurement of extravascular lung water (EVLW) may be useful in the treatment of critically ill children and can be performed at the bedside using the transpulmonary thermodilution technique (TPTD). There are currently no data to verify the accuracy of these measurements in (small) children. We compared the results of TPTD measurement with the clinical gold standard transpulmonary double indicator dilution (TPDD) measurement in young children.

Design: Prospective clinical study in children.

Setting: Catheterization laboratory of a university hospital.

Patients and methods: Twelve children (<2 yrs or <12 kg) under general anesthesia.

Interventions: None.

Measurements and main results: Measurements were performed using injections of ice-cold indicator (saline or dye) through a central venous catheter. Mean cardiac index was 3.91 L/min/m, mean intrathoracic blood volume index (ITBVITPDD) was 614.9 mL/m, and mean extravascular lung water index (EVLWITPDD) was 11.7 mL/kg. The correlation coefficient between EVLWITPDD and EVLWITPTD is 0.96 (95% confidence interval: 0.87-0.99; p < 0.0001). Bland-Altman analysis for EVLW measurements showed a mean bias of 2.34 mL/kg (18.13%) and limits of agreement +/-2.97 mL/kg (19.78%). The difference between measurements via the right atrium compared with the femoral vein was 2.8% for cardiac output, 8.2% for global end-diastolic volume, and 0.1% for EVLW.

Conclusion: Clinical measurement of EVLW in young children can be performed using the TPTD with the injection catheter inserted in the femoral vein. Further studies are needed to clarify the clinical value of these measurements.

PubMed Disclaimer

LinkOut - more resources