Reintroduction of Riva-Rocci measurements to determine systolic blood pressure?
- PMID: 19075314
Reintroduction of Riva-Rocci measurements to determine systolic blood pressure?
Erratum in
- Neth J Med. 2010 Jan;68(1):14. Bos, J-W [corrected to Bos, W J]
Abstract
Introduction: In 1896, Riva-Rocci introduced the upper arm cuff to measure systolic blood pressure. In 1905, Nicolai Sergeivich Korotkoff added the auscultatory technique, allowing measurement of both systolic and diastolic blood pressure. Both methods have, to our knowledge, never been formally tested against each other. In this study, we want to fill this gap in history.
Methods: We measured systolic blood pressure by the Korotkoff sound technique and approximated the Riva-Rocci technique by measuring cuff pressure at the moment that the first pulsation became visible in Finapres readings, at the finger. This proxy of the Riva-Rocci technique allows an objective, offline, analysis. Measurements were performed simultaneously on the same arm in 57 subjects.
Results: Systolic blood pressure measured by the Korotkoff sound technique was 167+/-30 mmHg (mean +/-SD ). Systolic blood pressure according to the Riva-Rocci technique was 165+/-32 mmHg. The Riva-Rocci technique underestimated measurements with the Korotkoff technique by 1.8+/-4.4 mmHg (NS, p=0.79).
Conclusion: Riva-Rocci measurements of systolic blood pressure may be as good as the traditionally used Korotkoff measurements.
Similar articles
-
The elevation of forearm arterial blood pressureduring Riva-Rocci-Korotkoff measurements.Blood Press Monit. 1996 Apr;1(2):141-147. Blood Press Monit. 1996. PMID: 10226216
-
Raising the arm before cuff inflation increases the loudness of Korotkoff sounds.Blood Press Monit. 2009 Dec;14(6):268-73. doi: 10.1097/MBP.0b013e3283330a67. Blood Press Monit. 2009. PMID: 19881335
-
Nexfin noninvasive continuous blood pressure validated against Riva-Rocci/Korotkoff.Am J Hypertens. 2009 Apr;22(4):378-83. doi: 10.1038/ajh.2008.368. Epub 2009 Jan 29. Am J Hypertens. 2009. PMID: 19180062
-
Accuracy of the blood pressure measurement.Minerva Cardioangiol. 2006 Aug;54(4):399-416. Minerva Cardioangiol. 2006. PMID: 17016412 Review.
-
Arterial pressure waveforms in hypertension.Minerva Med. 2003 Aug;94(4):229-50. Minerva Med. 2003. PMID: 14605588 Review.
Cited by
-
Advances in Cardiovascular Wearable Devices.Biosensors (Basel). 2024 Oct 30;14(11):525. doi: 10.3390/bios14110525. Biosensors (Basel). 2024. PMID: 39589984 Free PMC article. Review.
-
Comparison of telemetry and high-definition oscillometry for blood pressure measurements in conscious dogs: effects of torcetrapib.J Am Assoc Lab Anim Sci. 2010 Jul;49(4):464-71. J Am Assoc Lab Anim Sci. 2010. PMID: 20819394 Free PMC article.
-
The association between anthropometric parameters, the metabolic syndrome and microalbuminuria in black Africans: the SABPA study.Cardiovasc J Afr. 2010 May-Jun;21(3):148-52. doi: 10.5830/cvja-2010-025. Cardiovasc J Afr. 2010. PMID: 20532453 Free PMC article.
-
Comparison of systolic blood pressure values obtained by photoplethysmography and by Korotkoff sounds.Sensors (Basel). 2013 Oct 31;13(11):14797-812. doi: 10.3390/s131114797. Sensors (Basel). 2013. PMID: 24184918 Free PMC article.
-
Assessment of non-communicable disease risk factors, functional performance, and health-related quality of life in adults: a comparative analysis in low-resourced urban and rural areas of South Africa.BMC Public Health. 2024 Jun 12;24(1):1580. doi: 10.1186/s12889-024-18964-2. BMC Public Health. 2024. PMID: 38867182 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical
Miscellaneous