Tensile strength of cementing agents on the CeraOne system of dental prosthesis on implants
- PMID: 19077583
- DOI: 10.1097/ID.0b013e31818c4947
Tensile strength of cementing agents on the CeraOne system of dental prosthesis on implants
Abstract
Purpose: The aim of this in vitro study was to evaluate the tensile strength of titanium cylinders cemented on stainless steel abutment mock-ups by the Cerazone system.
Methods and materials: Four types of cements were used: glass ionomer, Fuji I (GC); zinc phosphate, Cimento LS (Vigodent); zinc oxide without eugenol, Rely x Temp NE (3M ESPE); and resin cement, Rely x ARC (3M ESPE). Four experimental groups were formed, each composed of 5 test specimens. Each test specimen consisted of a set of 1 cylinder and 1 stainless steel abutment mock-up. All cements tested were manipulated in accordance with manufacturers' instructions. A static load of 5 Newtons (N) for 2 minutes was used to standardize the procedure. The tensile tests were performed by a mechanical universal testing machine (EMIC DL500MF) at a crosshead speed of 0.5 mm/min.
Results: The highest bonding values resulting from the experiment were obtained by Cimento LS (21.86 MPa mean), followed by the resin cement Rely x ARC (12.95 MPa mean), Fuji I (6.89 MPa mean), and Rely x Temp NE (4.71 MPa mean). The results were subjected to analysis of variance (ANOVA) and the Student's t test.
Conclusion: The cements differed amongst them as regards tensile strength, with the highest bonding levels recorded with zinc phosphate (Cimento LS) and the lowest with the zinc oxide without eugenol (Rely x Temp NE).
Similar articles
-
The effects of abutment taper, length and cement type on resistance to dislodgement of cement-retained, implant-supported restorations.J Prosthodont. 2003 Jun;12(2):111-5. doi: 10.1016/S1059-941X(03)00006-8. J Prosthodont. 2003. PMID: 12964683
-
Retentiveness of various luting agents used with implant-supported prosthesis: an in vitro study.J Oral Implantol. 2014 Dec;40(6):649-54. doi: 10.1563/AAID-JOI-D-12-00008. J Oral Implantol. 2014. PMID: 25506659
-
Tensile bond strength of cast commercially pure titanium and cast gold-alloy posts and cores cemented with two luting agents.J Prosthet Dent. 2008 Feb;99(2):141-7. doi: 10.1016/S0022-3913(08)60030-X. J Prosthet Dent. 2008. PMID: 18262015
-
A clinically focused discussion of luting materials.Aust Dent J. 2011 Jun;56 Suppl 1:67-76. doi: 10.1111/j.1834-7819.2010.01297.x. Aust Dent J. 2011. PMID: 21564117 Review.
-
Dental cements.Curr Opin Dent. 1991 Apr;1(2):228-34. Curr Opin Dent. 1991. PMID: 1777671 Review.
Cited by
-
Retentiveness of implant-supported metal copings using different luting agents.Dent Res J (Isfahan). 2012 Jan;9(1):13-8. doi: 10.4103/1735-3327.92921. Dent Res J (Isfahan). 2012. PMID: 22363357 Free PMC article.
-
The influence of implant abutment surface roughness and the type of cement on retention of implant supported crowns.J Clin Diagn Res. 2015 Mar;9(3):ZC05-7. doi: 10.7860/JCDR/2015/12060.5621. Epub 2015 Mar 1. J Clin Diagn Res. 2015. PMID: 25954694 Free PMC article.
-
Retention of Implant Supported Metal Crowns Cemented with Different Luting Agents: A Comparative Invitro Study.J Clin Diagn Res. 2016 Apr;10(4):ZC61-4. doi: 10.7860/JCDR/2016/15912.7635. Epub 2016 Apr 1. J Clin Diagn Res. 2016. PMID: 27190954 Free PMC article.
-
Application of Semipermanent Cements and Conventional Cement with Modified Cementing Technique in Dental Implantology.Acta Stomatol Croat. 2021 Dec;55(4):367-379. doi: 10.15644/asc55/4/4. Acta Stomatol Croat. 2021. PMID: 35001932 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Miscellaneous