Comparison of radionuclide estimation of glomerular filtration rate using technetium 99m diethylenetriaminepentaacetic acid and chromium 51 ethylenediaminetetraacetic acid
- PMID: 1908780
- DOI: 10.1007/BF02258429
Comparison of radionuclide estimation of glomerular filtration rate using technetium 99m diethylenetriaminepentaacetic acid and chromium 51 ethylenediaminetetraacetic acid
Abstract
Simultaneous measurements of the clearance rates of technetium 99m diethylenetriaminepentaacetic acid (99mTc-DTPA) and chronium 51 ethylenediaminetetraacetic acid (51Cr-EDTA) were performed in 30 patients with a range of renal function (glomerular filtration rates between 9 and 120 ml/min). Using multiple blood samples, the two clearance values correlated well (r = 0.991, standard error 3.9 ml/min), but DTPA clearance was systematically higher by 7.6%. For each radiopharmaceutical, an equation was derived to correct clearance values obtained using only plasma samples taken at 2 and 4 h for the systematic error inherent in this technique compared with analysis of the complete plasma concentration-time curve. The root mean square error remaining after application of these equations was 1.9 ml/min for both the EDTA and DTPA data. The corresponding errors obtained using the equation derived by Brochner-Mortensen for EDTA plasma clearance were 2.2 ml/min and 1.9 ml/min, respectively, these values were not significantly different from those obtained using the equations derived in this study.