Accuracy and validity of space analysis and irregularity index measurements using digital models
- PMID: 19113071
Accuracy and validity of space analysis and irregularity index measurements using digital models
Abstract
Background: Digital copies of study models may avoid the storage and retrieval issues of plaster study models, but measurements made on digital models may not be as accurate as measurements made on traditional study models.
Aim: To determine the reliability and validity of tooth size-arch length discrepancies (TALD), irregularity indices and arch lengths (four- and six-segment analyses) measured directly on study models with digital calipers with the same measurements measured on digital copies of the study models with proprietary software.
Methods: The irregularity indices and TALDs (four- and six-segments) were measured on 50 sets of pretreatment plaster models. The plaster models were measured using manual calipers with a digital readout. The models were then couriered to OrthoCAD and digital copies emailed to the authors. The digital models were measured with the proprietary software provided with the digital models. Repeat measurements of the TALDs and the irregularity indices were subjected to intraclass correlations (ICC) to assess the reliability. The least squares means of variation was used to assess validity and the impact of measuring arch length (four- and six-segments) on the digital models, and the implications on the TALDs.
Results: There were high correlations (ICC) ranging from 98.6-99.9 per cent for both the irregularity indices and the TALDs. The choice of manual over computer and four-segment over six-segment analysis had a significant effect when measuring lower arch lengths (p < 0.05), but they had no effect on the upper arch findings.
Conclusions: Reliable measurements of the irregularity index and the TALD can be made on digital models. Computer measurements of TALDs on digital models were more consistent than manual measurements of TALDs on plaster models. Six-segment analyses of lower arch lengths on digital and plaster models gave more consistent findings than the four-segment analyses.
Similar articles
-
Comparison of space analysis evaluations with digital models and plaster dental casts.Am J Orthod Dentofacial Orthop. 2009 Jul;136(1):16.e1-4; discussion 16. doi: 10.1016/j.ajodo.2008.11.019. Am J Orthod Dentofacial Orthop. 2009. PMID: 19577140
-
Validity, reliability and reproducibility of three methods used to measure tooth widths for bolton analyses.Aust Orthod J. 2009 Nov;25(2):97-103. Aust Orthod J. 2009. PMID: 20043542
-
Accuracy of space analysis with emodels and plaster models.Am J Orthod Dentofacial Orthop. 2007 Sep;132(3):346-52. doi: 10.1016/j.ajodo.2005.08.044. Am J Orthod Dentofacial Orthop. 2007. PMID: 17826603
-
Assessment of the reliability of measurements taken on digital orthodontic models obtained from scans of plaster models in laboratory scanners. A systematic review and meta-analysis.Eur J Orthod. 2022 Sep 19;44(5):522-529. doi: 10.1093/ejo/cjac005. Eur J Orthod. 2022. PMID: 35363303
-
[Stereophotogrammetry and its use in stomatological research--a literature survey (author's transl)].Zahn Mund Kieferheilkd Zentralbl. 1980;68(8):786-800. Zahn Mund Kieferheilkd Zentralbl. 1980. PMID: 6452771 Review. German. No abstract available.
Cited by
-
Evaluation of the agreement of horizontal and vertical linear measurements obtained from digital models, printed models and direct measurements.Dental Press J Orthod. 2024 Oct 7;29(5):e242460. doi: 10.1590/2177-6709.29.5.e242460.oar. eCollection 2024. Dental Press J Orthod. 2024. PMID: 39383373 Free PMC article.
-
Overall and Anterior Tooth Size Ratios in a Group of Emiratis.Open Dent J. 2018 Sep 28;12:655-663. doi: 10.2174/1745017901814010655. eCollection 2018. Open Dent J. 2018. PMID: 30369975 Free PMC article.
-
Accuracy and reliability of measurements performed using two different software programs on digital models generated using laser and computed tomography plaster model scanners.Korean J Orthod. 2020 Jan;50(1):13-25. doi: 10.4041/kjod.2020.50.1.13. Epub 2020 Jan 22. Korean J Orthod. 2020. PMID: 32042716 Free PMC article.
-
Linear measurements using virtual study models.Angle Orthod. 2012 Nov;82(6):1098-106. doi: 10.2319/110311-681.1. Epub 2012 Apr 24. Angle Orthod. 2012. PMID: 22530811 Free PMC article.
-
Are there differences between comparison methods used to evaluate the accuracy and reliability of digital models?Dental Press J Orthod. 2017 Feb;22(1):65-74. doi: 10.1590/2177-6709.22.1.065-074.oar. Dental Press J Orthod. 2017. PMID: 28444012 Free PMC article.